Installing and Setting Up the Symmetricom 5087B
2
Symmetricom 5087B Operations and Maintenance Manual Rev. E
17
Adjusting the Output Level
You can adjust the output level to a value between 0.75 Vrms and 3
Vrms.
Required equipment:
•
Oscilloscope
•
Small flat-head screwdriver
To adjust the output level
1
Connect an oscilloscope to the output you want to change.
2
Using a small flat-head screwdriver, turn the potentiometer next to
the LED for the output until the oscilloscope shows the output level
you want.