Inspection Items
Acceptance Standard Qty Picture
Period
Action/Recomendation
17 Lower side guard (right & left)
- Side guard bolt, hex.nut M8,
plate washer M8
Loose/released,
sluggish/
stuck,broken, sound
when moving
4
6
months
Fasten, change if
broken, grease if
needed
18 Infuse holder & hangers
Tightening, bent,
broken
1&4
6
months
Fasten/adjust, repair,
change if broken/
fracture
19 Upper & lower pipe (section)
Bent, fracture
2
6
months
Fasten/adjust, repair,
change if fracture/
broken-if needed
20 Locking bolt
Crack, worn out/
broken, sluggish
1
3
months
Fasten, change if worn
out/broken, grease if
needed
If there are any problems, please see Manual Assembly, Part Catalog etc.
Read the manual book carefully first before using or operating this product.
Final disposal and decommissioning of medical devices (B3 & Non B3) shall be done using the disposal procedure and
withnesseth by the competent authority.