Operating Instructions
21
6.5 Menu structure
The menu structure supports the user when adjusting the measuring range, the calibration, the measurement
values and the choice of additional functions. In this connection, the numbering both on the display and in
the PC interface is identical:
1. Measurement range
1.1 Tag No
Input:
Free text (10 characters)
Name of the measurement point or product.
1.2 Unit
Input:
Unit text, e. g. kg
Required mass flow unit.
1.3 Time scale
Selection:
hour / minute /second time base for the integration by the
totaliser and the pulse output.
1.4 Decimal point
Selection:
0000, 0.000, 00.00, 000.0 Number representation and decimal point-
accuracy in the measurement menu.
1.5 Set point low
Input:
0 … 9999
Throughput rates under this value will
not be displayed at the current output.
This does not concern the display indicator,
totaliser or pulse output.
1.6 Set point hight
Input:
0 … 9999
Throughput rates above this value
will not be displayed at the current output.
This does not concern the display indicator,
totaliser or pulse output.
1.7 Filter
Input:
0.0 s … 999.9 s
Filtering of measurement for the indicator
and the output values.
1.8 Median filter
Input:
1 … 128
Distortion filter for peak measurement -
input. Acts like an additional filter to the filter
value [ s ] [ 1.7 ].
This filter directly impacts the raw value
extraction and thus all other calculation
steps.