MODEL 2200-3000 RETROFIT OPEN STREAM CURRENT METER / PAGE 5 of 7
Supplement to Model 3000 Instruction Manual
QUICK OPERATING INSTRUCTIONS FOR THE MODEL 2200-3000
1. Connect 2200 Sensor cable to the Model 3000 Indicator by use of the twist-lock connector.
2. Attach the Sensor to a meter using the knurled slip ring. Check that the bucket wheel rotates freely
and that the fiber-optics rotor does not contact the sensor or the adapter.
4. Choose the Prop (calibration number appropriate for the sensor-meter combination you have connected
to the Indicator). Normally it will be Prop 4 or 5. (4=AA, optic, 5=Pygmy,optic).
5. Place the Indicator in the Count mode and spin the meter to be sure that the sensor is producing pulses.
With the normal "two hole" fiber-optic rotor you should get two pulses per meter revolution. If
confirmed then:
6. Choose the operation desired by use of the key pad. (Count Revolutions, Velocity, or Discharge
Mode).
7. Place the meter in the stream and begin taking readings.
NOTE: A separate section of the instructions refers to the Model 3000 Indicator. In most cases references to
the 2200 portion of the instrument is either not necessary or is implied since the Model 3000 instrument was
designed to operate with many different kinds of sensors including the "2200" version for Price Type meters
with optic sensors.