Swoffer Instruments 2200-3000 Instruction Manual Download Page 5

MODEL 2200-3000  RETROFIT OPEN STREAM CURRENT METER / PAGE 5 of 7 

Supplement to Model 3000 Instruction Manual

 

 

QUICK OPERATING INSTRUCTIONS FOR THE MODEL 2200-3000 

1.  Connect 2200 Sensor cable to the Model 3000 Indicator by use of the twist-lock connector. 
2.  Attach the Sensor to a meter using the knurled slip ring.  Check that the bucket wheel rotates freely 

and that the fiber-optics rotor does not contact the sensor or the adapter. 

4.  Choose the Prop (calibration number appropriate for the sensor-meter combination you have connected 

to the Indicator). Normally it will be Prop 4 or 5. (4=AA, optic, 5=Pygmy,optic). 

5.  Place the Indicator in the Count mode and spin the meter to be sure that the sensor is producing pulses. 

With the normal "two hole" fiber-optic rotor you should get two pulses per meter revolution. If 
confirmed then: 

6.  Choose the operation desired by use of the key pad. (Count Revolutions, Velocity, or Discharge 

Mode). 

7.  Place the meter in the stream and begin taking readings. 

NOTE: A separate section of the instructions refers to the Model 3000 Indicator. In most cases references to 
the 2200 portion of the instrument is either not necessary or is implied since the Model 3000 instrument was 
designed to operate with many different kinds of sensors including the "2200" version for Price Type meters 
with optic sensors. 

SWOFFER INSTRUMENTS, INC. 

1112 S. 344

th

 St., Suite 302 

Federal Way, WA  98003  U.S.A. 

http://www.swoffer.com

 

Tel. (253) 661-8706 

Summary of Contents for 2200-3000

Page 1: ...red by the USGS Price Type Meters have been mechanically reliable for many years however accuracy has been a function of careful maintenance and adjustment of the delicate cats whisker contactors and...

Page 2: ...ALLATIONS refer to drawings in back AA METER 1 Remove contact chamber assembly entirely 2 Remove pivot shaft from yoke at lower end of the bucket wheel 3 Unscrew and remove upper rotor shaft from the...

Page 3: ...oles in the upper threads of the adapter Make sure the adapter is tight to the yoke otherwise removal of the sensor nut may unscrew the adapter if the threads are a bit snug 6 Re install the lower piv...

Page 4: ...he Model 3000 Indicator operating instructions for further information on calibration and calibration corrections AA METER RETROFIT PROBLEMS 1 In some AA Yokes it has been found that a deep chamfer ha...

Page 5: ...Indicator in the Count mode and spin the meter to be sure that the sensor is producing pulses With the normal two hole fiber optic rotor you should get two pulses per meter revolution If confirmed th...

Page 6: ...7 Supplement to Model 3000 Instruction Manual 2200 A Sensor A3 RETAINING NUT 011 O RING A14 FIBER OPTICS CARRIER ASSEMBLY A2 YOKE ADAPTER 250 10 32 R H M SCREW A4 SHAFT YOKE 3 32 05 allen wrench use...

Page 7: ...nt to Model 3000 Instruction Manual 2200 A Sensor Cable assembly A3 RETAINING NUT 011 O RING A14 FIBER OPTICS CARRIER ASSEMBLY A1 YOKE ADAPTER PYGMY A5 SHAFT 10 32 x 1 4 screw 3 32 05 Allen Wrench Use...