Swoffer Instruments 2200-3000 Instruction Manual Download Page 2

MODEL 2200-3000  RETROFIT OPEN STREAM CURRENT METER / PAGE 2 OF 7  

Supplement to Model 3000 Instruction Manual

 

 

MODEL 2200-3000 PARTS LIST 

PART NUMBER 

DESCRIPTION 

QUANTITY 

2200-A 

Sensor-cable assembly, 10' 

3000 Indicator 

Digital readout indicator 

2200-A1 

Sensor adapter, Pygmy meter 

2200-A2 

Sensor adapter, AA meter 

2200-A4 

Replacement Rotor Shaft, AA meter 

2200-A5 

Replacement Rotor Shaft, Pygmy meter 

2200-011 

"O" Ring seals, (2 spares also provided) 

2200-A14 

Fiber-Optic Carrier Rotors 

2200-A16 

Adapter protection Caps 

.05" Allen wrench 

Used as Rotor Installation Wrench 

10-32 X ¼" Screw 

Replaces cat-whisker contactor PYGMY meter only 

NECK STRAP 

Snap swivel for attachment to Indicator back 

CHEST-WAIST HARNESS  

For hands free use of the 3000 Indicator 

AA BATTERIES 

One set (4) required for operation plus a spare set 

CD-ROM 

IBM PC compatible only, file transfer software, etc. 

USB cable 

Interface cable for pc USB port-to-M3000 meter 

INSTRUCTIONS  

FOR OPERATION AND MAINTENANCE 

1 Set

 

SENSOR ADAPTER INSTALLATIONS (refer to drawings in back) 

AA METER 

1.  Remove contact chamber assembly entirely. 
2.  Remove pivot shaft from yoke at lower end of the 

bucket wheel.  

3.  Unscrew and remove upper rotor shaft from the 

bucketwheel and replace it with part 2200-A4.  
Carefully mate the threads of the new shaft to the 
bucketwheel threads and tighten shaft ONLY with the 
.05" Allen wrench inserted in the wrench hole in the 
shaft (to prevent damage to the bearing surfaces). 

4.  Install Sensor Adapter (2200-A2) through the top of 

the yoke, carefully inserting the upper portion of the 
new rotor shaft through the hole in the bottom of the 
adapter.  When seated, the AA adapter should be 
locked in place by tightening the front yoke screw.  

See AA Meter Retrofit Problems ahead for fit detail 
and possible problems with installation of this 
component.

 

5.  Re-install the lower (pivot) shaft and adjust vertical 

position of bucketwheel as necessary. 

6.  Using the brass blade driver supplied install the Fiber-Optic Carrier Rotor (2200-A14) through the top 

of the Adapter and onto the threads of the new rotor shaft.  DO NOT TIGHTEN TOO TIGHT TO 
AVOID STRIPPING THREADS.  The top of the Fiber-Optic Rotor must not be above the shoulder 
inside the Adapter.  See step 4. 

Summary of Contents for 2200-3000

Page 1: ...red by the USGS Price Type Meters have been mechanically reliable for many years however accuracy has been a function of careful maintenance and adjustment of the delicate cats whisker contactors and...

Page 2: ...ALLATIONS refer to drawings in back AA METER 1 Remove contact chamber assembly entirely 2 Remove pivot shaft from yoke at lower end of the bucket wheel 3 Unscrew and remove upper rotor shaft from the...

Page 3: ...oles in the upper threads of the adapter Make sure the adapter is tight to the yoke otherwise removal of the sensor nut may unscrew the adapter if the threads are a bit snug 6 Re install the lower piv...

Page 4: ...he Model 3000 Indicator operating instructions for further information on calibration and calibration corrections AA METER RETROFIT PROBLEMS 1 In some AA Yokes it has been found that a deep chamfer ha...

Page 5: ...Indicator in the Count mode and spin the meter to be sure that the sensor is producing pulses With the normal two hole fiber optic rotor you should get two pulses per meter revolution If confirmed th...

Page 6: ...7 Supplement to Model 3000 Instruction Manual 2200 A Sensor A3 RETAINING NUT 011 O RING A14 FIBER OPTICS CARRIER ASSEMBLY A2 YOKE ADAPTER 250 10 32 R H M SCREW A4 SHAFT YOKE 3 32 05 allen wrench use...

Page 7: ...nt to Model 3000 Instruction Manual 2200 A Sensor Cable assembly A3 RETAINING NUT 011 O RING A14 FIBER OPTICS CARRIER ASSEMBLY A1 YOKE ADAPTER PYGMY A5 SHAFT 10 32 x 1 4 screw 3 32 05 Allen Wrench Use...

Reviews: