MODEL 2200-3000 RETROFIT OPEN STREAM CURRENT METER / PAGE 2 OF 7
Supplement to Model 3000 Instruction Manual
MODEL 2200-3000 PARTS LIST
PART NUMBER
DESCRIPTION
QUANTITY
2200-A
Sensor-cable assembly, 10'
1
3000 Indicator
Digital readout indicator
1
2200-A1
Sensor adapter, Pygmy meter
1
2200-A2
Sensor adapter, AA meter
1
2200-A4
Replacement Rotor Shaft, AA meter
1
2200-A5
Replacement Rotor Shaft, Pygmy meter
1
2200-011
"O" Ring seals, (2 spares also provided)
4
2200-A14
Fiber-Optic Carrier Rotors
2
2200-A16
Adapter protection Caps
2
.05" Allen wrench
Used as Rotor Installation Wrench
1
10-32 X ¼" Screw
Replaces cat-whisker contactor PYGMY meter only
1
NECK STRAP
Snap swivel for attachment to Indicator back
1
CHEST-WAIST HARNESS
For hands free use of the 3000 Indicator
1
AA BATTERIES
One set (4) required for operation plus a spare set
2
CD-ROM
IBM PC compatible only, file transfer software, etc.
1
USB cable
Interface cable for pc USB port-to-M3000 meter
1
INSTRUCTIONS
FOR OPERATION AND MAINTENANCE
1 Set
SENSOR ADAPTER INSTALLATIONS (refer to drawings in back)
AA METER
1. Remove contact chamber assembly entirely.
2. Remove pivot shaft from yoke at lower end of the
bucket wheel.
3. Unscrew and remove upper rotor shaft from the
bucketwheel and replace it with part 2200-A4.
Carefully mate the threads of the new shaft to the
bucketwheel threads and tighten shaft ONLY with the
.05" Allen wrench inserted in the wrench hole in the
shaft (to prevent damage to the bearing surfaces).
4. Install Sensor Adapter (2200-A2) through the top of
the yoke, carefully inserting the upper portion of the
new rotor shaft through the hole in the bottom of the
adapter. When seated, the AA adapter should be
locked in place by tightening the front yoke screw.
See AA Meter Retrofit Problems ahead for fit detail
and possible problems with installation of this
component.
5. Re-install the lower (pivot) shaft and adjust vertical
position of bucketwheel as necessary.
6. Using the brass blade driver supplied install the Fiber-Optic Carrier Rotor (2200-A14) through the top
of the Adapter and onto the threads of the new rotor shaft. DO NOT TIGHTEN TOO TIGHT TO
AVOID STRIPPING THREADS. The top of the Fiber-Optic Rotor must not be above the shoulder
inside the Adapter. See step 4.