TENSION CONTROL OF FOCUSING MOVEMENT 

 

The focus tension is easily adjusted by using the Swift adjustment 
wrench.  This wrench fits the tension collar found on the focusing 

controls, between the knob and upright support.  A clockwise turn 
of this collar moves it toward the upright support and increases 

tension, while a counter clockwise turn moves the collar toward the 

knob and decreases tension (see illustration). 
 

 

 

 

 

 

 

 

COMMON PROBLEMS IN MICROSCOPY 

If you have a problem, you may be able to correct it yourself. Here 
are a few common problems and easy solutions you may want to try 
before calling for service. 
 
CAUTION – Never disassemble mechanical or optical components. 
This servicing should only be done by an authorized Swift technician. 
The Limited Lifetime Warranty will be null and void if the 
mechanical or optical components are disassembled by a non-Swift 
dealer. 

 

A. PROBLEM – The illuminator light does not come on 

CORRECTION –  

1. Make sure the microscope is plugged into a functional 

electrical socket. 
2. Make sure the bottom illuminator dimmer control is set above 

“0”  
3. The bulb may need to be replaced.  See “Bulb replacement” 

on page 8. 