Swift M27LED Series Use And Care Manual Download Page 5

 

To change the top LED illuminator: 
1.  Unscrew the top illuminator housing cover that also holds the light  
 

filter in place. 

 
2.  Remove the metal ring spacer and pull the LED bulb and round    
      circuit card assembly out of the illuminator housing. 
 
3. 

Turn the LED/circuit card over and unplug the connector to separate  

      the wires  from the assembly. 

 
4.

 

Reverse the previous steps to install the new LED bulb. 

 
 
To change the bottom LED illuminator: 
1.  Push the stage clips to the side of the stage plate and remove the 
 

stage plate that is covering the bottom illuminator. 

 
2.  Pull the LED bulb out of the bulb socket. 
 
3.  Insert the metal pins of the new bulb into the bulb socket. 
 
4.  Reverse the previous steps to install the new LED bulb. 
 
Note:  If the new bulb does not work after installation, the polarity of 
the bulb may have been reversed when installing the new bulb. Simply 
unplug the new bulb from the bulb socket and turn the bulb to insert 
the bulb’s metal pins into the opposite bulb housing connections.  
 
FOCUS TENSION ADJUSTMENT: 
The focus tension adjustment is easily adjusted by using the C-Wrench 
(Part MT202). This wrench fits the tension collar found on the focusing 
control spindle, between the adjustment knob and body of the 
microscope. A clockwise turn of this collar moves it toward the upright 
support and increases tension, while a counter-clockwise turn moves 
the collar toward the knob and decreases tension. 

Summary of Contents for M27LED Series

Page 1: ...1 Swift M27LED Series Stereo Microscope Use and Care Manual SWIFT OPTICAL Enduring Quality and Technical Excellence...

Page 2: ...able field of view The M27LED incorporates features to enhance and provide maximum versatility of the instrument for classroom use It is built to rigid optical and mechanical standards The scope is th...

Page 3: ...the clutch system will slip to prevent damage to mechanical parts ILLUMINATION The top incident and bottom trans base illuminators are controlled by two on off switches located on the base of the mic...

Page 4: ...l system is now adjusted to your particular vision 5 Turn the turret to a higher power for increased magnification SERVICE Always unplug the microscope from the power outlet while performing service a...

Page 5: ...nsert the metal pins of the new bulb into the bulb socket 4 Reverse the previous steps to install the new LED bulb Note If the new bulb does not work after installation the polarity of the bulb may ha...

Page 6: ...urfaces should be cleaned frequently with mild detergent and a soft cloth TROUBLE SHOOTING A PROBLEM The image does not stay in focus CORRECTION Use Swift C Wrench MT202 to tighten the tension collar...

Page 7: ...NA 4X NA 48 mm Depth of View 0 17 1 05mm 0 13 1 05mm Dimensions 10 L X 5 W X 12 5 H Weight 6 6 lbs Illumination Top Incident LED 3 4V 0 3W Bottom Trans base LED 3 4V 0 06W PARTS AND ACCESSORIES MA147...

Page 8: ...FT OPTICAL INSTRUMENTS INC LIMITED LIFETIME WARRANTY Please see our website www swiftoptical com for complete warranty details and exclusions Swift Optical Instruments Inc 877 967 9438 www swiftoptica...

Reviews: