the small allen wrench (.09mm) that was included with the microscope 
to loosen the set screws that hold the black plastic illuminator housing 
onto the base of the microscope. Remove the illuminator housing to 
expose the LED. Simply pull the LED straight up to remove it from the 
light socket. Align the 2 metal socket pins with the holes at the bottom 
of the new LED and push the LED onto the socket. Re-install the 
illuminator housing. (If the new LED does not work, try repeating the 
above procedure. Remove the bulb and re-install it using the opposite 
bulb socket pins to reverse the polarity.) 
 

 

TENSION CONTROL OF FOCUSING MOVEMENT 

 
The focus tension is easily adjusted by using the collar found on the 
coarse focusing controls, between the knob and upright support. A 
clockwise turn of this collar moves it toward the upright support and 
increases tension, while a counter clockwise turn moves the collar 
toward the knob and decreases tension. 

 

 

COMMON PROBLEMS IN MICROSCOPY 

 

A.  PROBLEM – The illuminator light does not come on.   

 

 

 

CORRECTION 

 

1.  Make sure the microscope is plugged into a functional electrical 

 

 

socket. 

 

2.  Make sure the batteries are properly charged. Plug the power 

    adapter in and see if the microscope will turn on (M2251C). 

  3.  The bulb may need to be replaced. See “Bulb Replacement on 
   

page 7.                                                     

 

B.  PROBLEM – Image appears “washed out” or weak. 

 

 

 

CORRECTION 

 

1.  Rotate disc diaphragm to smaller aperture or slightly close the 

 

 

iris diaphragm. 

 

2.  Objective lens is dirty. Clean as described under “Cleaning”. 

 

3.  Eyepiece is dirty. Clean as described under “Cleaning”. 

Summary of Contents for M2250 Series

Page 1: ...Swift M2250 Series Microscope User and Care Manual SWIFT OPTICAL Enduring Quality and Technical Excellence...

Page 2: ...2...

Page 3: ...ocked on eyepieces objectives and stage clips and a slip clutch focusing system to protect gears from over focusing The M2250 Series is offered with several illumination options standard tungsten extr...

Page 4: ...s diaphragm M2251C M2251CL The function of the condenser is to provide full illumination to the specimen plane and to enhance the resolution and contrast of the object being viewed DISC DIAPHRAGM M225...

Page 5: ...ENS in attempting to transmit light through glass much of the light is lost through reflection Coating a lens increases the light transmission by reducing or eliminating reflection thus allowing more...

Page 6: ...ld be rotated to the next smaller aperture of the disc diaphragm or the iris diaphragm should be closed slightly This will increase contrast in the specimen s image 6 Move the slide to place the speci...

Page 7: ...14281 Power adapter for M2251C 4 5V 1A MA14283 Cold Holder for M2251B or M2251CL Pair MA2201 120V 20W bulb for M2251B MA2215 3 4V 06W LED for M2251C M2251CL MA268 Stage Clips MA336 Dustcover CARE OF Y...

Page 8: ...containing chemicals which may corrode your microscope Also be sure that the objectives are placed in the lowest possible position and the rotating head is turned inward and not protruding from the b...

Page 9: ...using the collar found on the coarse focusing controls between the knob and upright support A clockwise turn of this collar moves it toward the upright support and increases tension while a counter cl...

Page 10: ...not be centered properly on the stage and needs to be re positioned to be in the optical path E PROBLEM Image of the specimen goes out of the focus all by itself CORRECTION Use the tension control co...

Page 11: ...11...

Page 12: ...12 Swift Optical Instruments Inc 877 967 9438 www swiftoptical com...

Reviews: