Swema SwemaCO2 Operating Instructions Manual Download Page 5

4.

4,11V, 50,4 hours of logging 

with chosen interval.

Pump OFF

LOG Int.S

Log interval:

 Adjust the log interval in seconds.  

Choose between 

16

, 30, 60, 120, 

240

, 600, 900, 

1200

, 1800, 

3600

7200

 second intervals. Use 

the 

Up 

and

 Down arrows

. The 

bold

 second inter-

vals are in phase with the CO

2

-sensors measuring 

frequency. If a logging is initiated without *, the 

logged measurements with these time constants 

will be measured values (non predicted - *).

Battery symbol - initiate logging

The display to the right shows the battery vol-

tage (V) and how long the instrument can log 

with the set log interval (hours until the memo-

ry is full). The battery life may be shorter. 

By pressing Enter a logging is initiated. Start 

once the star disappears to collect a measured 

value. Longer log intervals can be chosen for 

longtime logging, see (LOG Int.S). Use a USB 

power supply to prevent power loss. See also 

6. 

Logging

.

PuP.

Pump:

 Use the 

Up 

and

 Down arrows

, to switch 

between 

ON/OFF. 

It  is  also  possible  to  switch 

the pump on and off during a measurement. The 

pump  absorbs  air  and  blows  it  into  the  Carbon 

dioxide sensor. Refrain from breathing on the in-

strument as this may affect the CO

2

-measurement 

even when the pump is switched on.

noti Fy

Notify:

 

ON

 or 

OFF

. The instrument will beep once for every thou-

sandth PPM. E.g. at 5300 ppm the instrument shall beep five times. 

If the ppm rises, the beeps will sound in higher pitched tones, and 

the backlight will turn on for five seconds. If the PPM is stable on the 

same thousandth, the beeps will sound in the same pitch. If the PPM 

drops, the beeps will also drop in pitch. The instrument waits for a 

few minutes between every beep-sequence. 

AdJ. 410 

Adjust

: Adjust the CO

2

-sensor to show 410 ppm. Enable the pump 

and absorb some fresh air. This is assumed to have 410 ppm CO

2.

This adjustment is done to prevent drift (offset) of the sensor.

Log interval

Summary of Contents for SwemaCO2

Page 1: ...wema swema se Web page www swema com Contents 1 Introduction 1 2 Start of instrument 2 3 Measurement Mode 2 4 Function keys 3 5 Settings 3 6 Logging 5 7 PC transfer 5 8 Saved data 7 9 Technical data 8...

Page 2: ...sh whilst also ensuring the fastest possible response time The instrument can display the average value as well as saving and logging values Measurement series can also be transferred to a PC using th...

Page 3: ...n shows the current measurement in real time Average Max Min The Left arrow average Up arrow max or Down arrow min can be held down respectively to check the average max or min value of a measurement...

Page 4: ...s 60min 30min 10min 5min and 1min OFF disables the auto shut off completely Unlt SySt Unit system Syst me international SI Uni ted States customary US change of international measurement SI or america...

Page 5: ...power supply to prevent power loss See also 6 Logging PuP Pump Use the Up and Down arrows to switch between ON OFF It is also possible to switch the pump on and off during a measurement The pump absor...

Page 6: ...To stop and save a log press All For a brief moment the display will read Notebook with n3 blinking which in this case means that the log was saved as measurement series 3 in notebook The instrument...

Page 7: ...down Erase saved data Erase makes it possible to delete saved measurement values Upon pressing Erase the text Erase is shown on the display to erase the entire memory press All Press Esc to cancel It...

Page 8: ...terval CO2 16 seconds for the measured value Two prognostications in between indicated by a on display In stable conditions prognostications equal measured value Barometer Twice per second Log Interva...

Reviews: