A-96.250.211 / 171121
AMI Sodium P
Program List and Explanations
95
2.2.2.5
Cal. History:
Shows the diagnostic values of the last pH electrode
calibrations, offset in mV and slope in mV/pH.
2.2.3
Miscellaneous
:
2.2.3.1
Case Temp:
Temperature in °C inside the transmitter.
2.3 Sample
2.3.1
Sample ID
: Shows the identification assigned to a sample. This
identification is defined by the user to identify the location of the
sample.
Temperature
: Actual temperature in °C and in Ohm
(NT5K).
Sample Flow
: Shows the actual sample flow B/s (bubbles per
second). Sample flow must be above 5 B/s.
(Raw value) (Hz):
Shows the raw value in Hz.
2.4 I/O State
Shows current status of all in- and outputs.
2.4.1
2.4.2
2.5 Interface
Only available if optional interface is installed.
Shows the programmed communication settings.
3 Maintenance
3.1 Calibration
In this menu, you can correct measured values, or calibrate offset
and slope of pH electrode.
3.1.1
Standard Sodium:
Possibility to correct the sodium sensor value.
Follow the commands on the screen. Save the value with the <En
-
ter> key.
See
, for more details.
3.1.2
Process pH:
Correction of pH electrode. See
, for details.
Typical offset of pH Electrode
0 mV
30 mV
Typical slope of pH Electrode
59 mV/pH
3 mV/pH
Alarm Relay
:
Active or inactive
Relay 1 and 2
:
Active or inactive
Input
:
Open or closed
Signal Output 1 and 2
: Actual current in mA
Signal Output 3
:
Actual current in mA (if option is installed)