AMI Deltacon Power
Installation
18
3.2. Mounting of Instrument Panel
The first part of this chapter describes the preparing and placing of
the system for use.
The instrument must only be installed by trained personnel.
Mount the instrument in vertical position.
For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation ma
-
terial is available:
– 4 Screws 6x60 mm
– 4 Dowels
– 4 Washers 6.4/12 mm
Mounting re
-
quirements
The instrument is only intended for indoor installation.
For dimensions see
3.3. Connecting Sample Inlet and Outlet
3.3.1
Swagelok Fitting Stainless Steel at Sample Inlet
Preparation
Cut the tube to length and deburr it. The tube must be straight and
free from blemishes for approximately 1,5 x tube diameter from the
end.
Lubrication with lubricating oil, MoS2, Teflon etc. is recommended for
the assembly and reassembly of bigger sized unions (thread, com
-
pression cone).
Installation
1
Insert the compression ferrule [C] and the compression cone [D]
into the union nut [B].
2
Screw on the union nut onto the body, do not tighten it.
3
Push the stainless steel pipe through the union nut as far as it
reaches the stop of the body.
4
Mark the union nut at 6 o’clock position.
5
While holding the fitting body steady, tighten the nut union 1¼ ro
-
tation using an open ended spanner.