5.3.1.1.4.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E034 is displayed in the message
list.
Range: 0–14 pH
5.3.1.1.4.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu
-
ates around the alarm value.
Range: 0–14 pH
5.3.1.1.4.45
Delay:
Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.1.5
Ammonia (if Calculations = yes)
5.3.1.1.5.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E035, is displayed in the message
list.
Range: 0–500 ppm
5.3.1.1.5.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E036 is displayed in the message
list.
Range: 0–500 ppm
5.3.1.1.5.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu
-
ates around the alarm value.
Range: 0–500 ppm
5.3.1.1.5.45
Delay:
Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.2
Sample Temp.
5.3.1.2.1
Temp. 1
5.3.1.2.1.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E007, is displayed in the message
list.
Range: 30–200 °C
5.3.1.2.1.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C