AMI Deltacon DG
Program List and Explanations
62
5.1.2.3.5
Temp. comp
:
5.1.2.2.3.1
Comp
.: Available compensation models:
Strong acids
5.1.2.4
Degasser
5.1.2.4.1
Mode: on, off, input
On
: The degasser is switched on.
Off
: The degasser is switched off.
Input
: The degasser can be switched on or off via signal input.
Note:
When the sample flow falls below 5 l/h, the degasser will
be switched off automatically.
5.1.2.4.2
Boiling point interval:
Defines the time interval of the boiling point de
-
termination. In most applications, the preset interval of six hours is
short enough to keep the carbon dioxide removal rate constant. In
water with an elevated carbon dioxide content (CC> 1 µS/cm) a
shorter boiling point interval is recommended.
Range: 1 h, 2 h, 4 h, 6 h
5.2 Signal Outputs
Note:
The navigation in the menu <Signal Output 1> and <Signal
Output 2> is equal. For reason of simplicity only the menu
numbers of Signal Output 1 are used in the following.
5.2.1
Signal Output 1:
Assign process value, the current loop range and a
function to each signal output.
5.2.1.1
Parameter:
Assign one of the process values to the signal output.
Available values:
Cond 1 (cc)
Cond 2 (sc)
Cond 3 (dc)
Temp. 1
Temp. 2
Temp. 3
Difference
Sample flow
pH
Ammonia