AMI Codes-II
Maintenance
A-96.250.571 / 050517
43
6.2.
Stop of Operation for Maintenance
1
Put the suction lances into a bucket with clean water.
2
Start <Fill system>.
The reagent tubes are flushed with water.
3
Remove the suction lances from the water.
4
Start <Fill system> again.
The water will be pumped out of the reagent tubes.
5
Stop sample flow.
6
Wait until level in flow cell is empty.
7
Shut off power of the instrument.
If pH option is installed:
8
Remove the pH sensor [A] from the flow cell.
9
Fill the protective cap [B] with water.
10
Put the protective cap onto the sensor tip.
A
B
pH sensor
Protective cap
A
B