In accordance with best measurement practices and ISO 1996-2 if measurements 
take place over longer periods of time, e.g. over a day or more then the meas-
urement system should be checked either acoustically or electrically at regular 
intervals, e.g. once or twice a day.
Following the ISO 1996-2 recommendations the SV 307 has been equipped with 
an 

Acoustic System Check (ASC)

 that can be activated remotely; either auto-

matically or upon manual request. The ASC is located in the top of the tube direct-
ly above the microphone of the SV 307.

System Check  

The ASC is based on a small speaker and the three MEMS microphones in the
SV 307. Activation of the ASC enables the speaker and triggers an alarm if any of 
the three MEMS microphones reports a different noise level than the others.

Remote System Check

System check configuration is available in SvanNET Configuration tab. Simply go 
to SvanNET / Configuration / Calibration - Automatic System Check to schedule 
the system check time. 

System Check in SvanNET APP 

To perform a manual system check simply open SvanNET application, select your 
SV 307 station and click the System Check button. 

Summary of Contents for SV 307

Page 1: ...is capable of logging sum mary results SR and spectral data with interval steps down to 1 second Also time history results TH down to 100 milliseconds Summary results SR include LEQ PEAK MAX MIN with...

Page 2: ...d under the bottom cover of the SV 307 The bottom cover operates as a gasket therefore it is neces sary to carefully unscrew the 4 screws to remove it It is strongly recommended to configure the remot...

Page 3: ...n Access Point Name SvanNET Application SvanNET Application is a simple software user interface used for the configura tion of the SV 307 to connect with the SvanNET service To obtain the application...

Page 4: ...alarm email The configuration of Status alarms ins available in the Status tab in SvanNET Projects SvanNET Projects is an optional payable extension to the basic service offering fully automated mana...

Page 5: ...SC is located in the top of the tube direct ly above the microphone of the SV 307 System Check The ASC is based on a small speaker and the three MEMS microphones in the SV 307 Activation of the ASC en...

Page 6: ...ure continuous operation of SV 307 Use of External Battery Optionally the SB 275 external battery 33 Ah can be used to extend the autono my of SV 307 The SB 275 is simply connected to DC IN socket in...

Page 7: ...service office at the following number or contact via the SVANTEK s website Service Office 48 22 51 88 320 or 48 22 51 88 322 Office hours are 9 00 a m to 5 00 p m Central European Time E mail at offi...

Reviews: