LAMBDA DG-4 PLUS & DG-5 PLUS OPERATION MANUAL – REV. 3.05B (20131009)
37
Table 4-3. DG-4-Only filter position movement commands.
Move to
Nominal
Filter
Position
Value
(Decimal;
hexadecimal; binary)
Move on
…
Atte-
nuation
Physical
Filter
Position
Notes
0
0, 00, 00000000
Command
0% 0
Light is blocked (“Shutter Closed”)
1
1; 01; 00000001
Command 100% 1
1
st
filter (25mm) at full
2
2; 02; 00000010
Command 100% 2
2
nd
filter (25mm) at full
3
3; 03; 00000011
Command 100% 3
3
rd
filter (25mm) at full
4
4; 04; 00000100
Command 100% 4
4
th
filter (25mm) at full
5
5; 05; 00000101
Command 50% 1
1
st
filter (25mm) at half
6
6; 06; 00000110
Command
50% 2
2
nd
filter (25mm) at half
7
7; 07; 00000111
Command
50% 3
3
rd
filter (25mm) at half
8
8; 08; 00001000
Command
50% 4
4
th
filter (25mm) at half
9
9; 09; 00001001
Command 33% 1
1
st
filter (25mm) at 1/3
10
10; 0A; 00001010
Command
33% 2
2
nd
filter (25mm) at 1/3
11
11; 0B; 00001011
Command
33% 3
3
rd
filter (25mm) at 1/3
12
12; 0C; 00001100
Command
33% 4
4
th
filter (25mm) at 1/3
13
13; 0D; 00001101
(Unused)
14
14; 0E; 00001110
(Unused)
15
15; 0F; 00001111
(Unused)
16
16; 10; 00010000
Trigger 0% 0 Light is blocked (“Shutter Closed”)
17
17; 11; 00010001
Trigger 100% 1 1
st
filter (25mm) at full
18
18; 12; 00010010
Trigger 100% 2 2
nd
filter (25mm) at full
19
19; 13; 00010011
Trigger 100% 3 3
rd
filter (25mm) at full
20
20; 14; 00010100
Trigger 100% 4 4
th
filter (25mm) at full
21
21; 15; 00010101
Trigger 50% 1 1
st
filter (25mm) at half
22
22; 16; 00010110
Trigger 50% 2 2
nd
filter (25mm) at half
23
23; 17; 00010111
Trigger 50% 3 3
rd
filter (25mm) at half
24
24; 18; 00011000
Trigger 50% 4 4
th
filter (25mm) at half
25
25; 19; 00011001
Trigger 33% 1 1
st
filter (25mm) at 1/3
26
26; 1A; 00011010
Trigger 33% 2 2
nd
filter (25mm) at 1/3
27
27; 1B; 00011011
Trigger 33% 3 3
rd
filter (25mm) at 1/3
28
28; 1C; 00011100
Trigger 33% 4 4
th
filter (25mm) at 1/3
29
29; 1D; 00011101
(Unused)
30
30; 1E; 00011110
(Unused)
31
31; 1F; 00011111
(Unused)
NOTE: “Move on Command” means move the light beam immediately upon receipt of the
external command. “Move on Trigger” means move the light beam only until triggered via
Strobe, Synch, or Synch gated to Strobe.