LAMBDA DG-4 PLUS & DG-5 PLUS OPERATION MANUAL – REV. 3.05B (20131009)
20
When the ENTER key is then pressed, FILTER NUMBER 8 is then displayed correctly.
Method 2: The main menu will look like this if the initial FILTER NUMBER = 10
The numeric key 0 is pressed first.
The numeric key 8 is pressed next with no further action required. FILTER NUMBER 8 is
then displayed correctly.
3.7.4
Move on Sync Pulse Mode
[MODE/4]
If the Move on Sync Pulse Mode is active, the controller will advance to the next filter in a
previously programmed FILTER NUMBER sequence (Ring Buffer) when an internal video
synchronization (sync) circuit detects the vertical sync pulse of a composite video signal
applied to the SYNC input on the rear of the Lambda DG-4. In this way, filter changes can be
coordinated to occur with the capture of video “frames”. The FILTER NUMBER sequence
can be programmed at your convenience and then executed later with precise timing (see
SET UP THE RING BUFFER). The FILTER NUMBERS can also be sent individually via
the serial or parallel port before the trigger event. This operational mode is selected by
pressing the MODE key followed by the 4 key to get to the MOVE ON SYNC PULSE menu.
Alternatively, you can “page” through the mode selection menus by pressing MODE followed
by the 1 key. The following menu will be displayed:
80 = FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND
8 = FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND
10 = FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND
00 = FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND
8 = FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND