SUTO S431 OEM Instruction And Operation Manual Download Page 20

 8 Installation

 8.5  Performing the electrical connection

The S431 has a 6-pole M12 connector for power supply and signal 
output.

 

6-pole M12 cable with open ends included 

Female (View from front)

Pin assignment connector plug M12 (6-pole)

Output

Pin 1

Pin 2

Pin 3

Pin 4

Pin 5

Pin 6

Analog

-I

isolated

-VB

+VB

SW

SW

+I

isolated

Modbus

GND

M

-VB

+VB

D+

D-

N/A

Wire color

Blue

White

Red

Yellow

Green

Black

Legend for pin assignment

GND

M

Ground for Modbus/RTU

-VB

Negative supply voltage

+VB

Positive supply voltage

+I

isolated

Positive 4 … 20 mA signal (isolated)

-I

isolated

Negative 4 ... 20 mA signal (isolated)

SW

Isolated pulse output (switch)

D+

Modbus/RTU data +

D-

Modbus/RTU data -

N/A

Not applicable

ATTENTION!
Do not screw the M12 connector using force. 
Otherwise it might damage the connection pins.

20

S431 OEM

Summary of Contents for S431 OEM

Page 1: ...English Instruction and Operation Manual S431 OEM Pitot Tube Compressor Flow Meter Inline...

Page 2: ...mpliance with this manual Should the device be tampered with in any manner other than a procedure which is described and specified in the manual the warranty is void and the manufacturer is exempt fro...

Page 3: ...14 8 2 Inlet and utlet sections 14 8 3 Welding Nipple and Welding Fixture 16 8 4 Removing the sensor 19 8 5 Performing the electrical connection 20 9 Configuration 21 9 1 Registering 21 9 2 Using Ser...

Page 4: ...manufacturer WARNING Compressed air Any contact with quickly escaping air or bursting parts of the compressed air system can lead to serious injuries or even death Do not exceed the maximum permitted...

Page 5: ...ons It is not allowed to use the product in explosive areas Please observe the national regulations before during installation and operation Remarks It is not allowed to disassemble the product Always...

Page 6: ...packaging which comes with the sensor Please make sure that the storage temperature of the sensor is between 30 70 C Avoid direct UV and solar radiation during storage For the storage the humidity mu...

Page 7: ...rsuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate r...

Page 8: ...low m3 h Pressure bar Total consumption m3 Temperature C 5 Features Flow and consumption measurement in wet or high mass flow or velocity applications Measurement at the compressor outlet Installation...

Page 9: ...of reading Sensor Differential pressure sensor with pitot tube Sampling rate 3 sec Turndown ratio 10 1 Response time t90 2 sec Consumption Selectable units m ft3 t lb l kg Pressure Accuracy 0 5 FS Sel...

Page 10: ...mA 1 5 A peaks for 3 sec 6 3 General data Configuration Wireless S4C FS App for mobile phones Material Housing PC ABS Wetted parts Stainless steel 1 4404 SUS 316L Miscellaneous Electrical connection 1...

Page 11: ...Ambient temperature 20 90 C Ambient humidity 95 rH Storage temperature 30 70 C Transport temperature 10 70 C Pipe sizes DN50 The specified accuracy is valid only within the minimum and maximum flows...

Page 12: ...8 32 183 2 647 6 470 8 200 1 961 19 611 32 69 326 9 1 154 11 543 10 250 3 064 30 642 50 07 510 7 1 804 18 035 12 300 4 412 44 125 73 54 735 4 2 597 25 971 Remark Flow range for Air at 6 bar g 50 C and...

Page 13: ...7 Dimensional drawing 7 Dimensional drawing S431 OEM 13...

Page 14: ...ed separately 8 1 Installation Options There are 2 installation options Installation inside compressors This installation requires a calibration of the S431 together with the compressor in the compres...

Page 15: ...ions can cause counter flow turbulence as well as turbulence in the direction of the flow The sensor is for indoor use only At an outdoor installation the sensor must be protected from solar radiation...

Page 16: ...horizontal pipe The diameter of the hole is 21 mm for the installation in a vertical pipe Then a welding nipple will be welded onto the pipe There are 4 types available Order No Description A4310 Weld...

Page 17: ...ation Horizontal pipe installation S695 4310 S4695 4310 horizontal pipe installation Horizontal pipe installation A4319 A4319 Horizontal pipe installation option Valid installation angle from 20 90 S4...

Page 18: ...S695 4311 In addition you also can use the customized welding fixture for welding processing to get the best positioning and welding quality as shown below Welding fixture in a horizontal pipe Weldin...

Page 19: ...not carefully follow the steps described below For your information The sensor is exposed to a force of 18 kg at the 8 bar g system pressure a force of 32 kg at 16 bar g system pressure Hold the senso...

Page 20: ...ted VB VB SW SW Iisolated Modbus GNDM VB VB D D N A Wire color Blue White Red Yellow Green Black Legend for pin assignment GNDM Ground for Modbus RTU VB Negative supply voltage VB Positive supply volt...

Page 21: ...lation of the S4C FS are as follows 1 On your Android device download the S4C FS app from Google Play Store or from the SUTO website 2 Install S4C FS 9 1 Registering Before using the S4C FS App you mu...

Page 22: ...e S431 As a registered user you can access the sensor through a wireless connection immediately If you are not a registered user you need to scan the QR code on the calibration certificate Scan QR Cod...

Page 23: ...ual Flow or FAD Free Air Delivery Gas constant It is set through the selected gas type Each gas type has its own specific gas constants When the gas type is other you have to input a gas constant Sens...

Page 24: ...9 Configuration 9 2 2 Unit Settings Select the desired physical unit for the different measurement channels 24 S431 OEM...

Page 25: ...or VSD compressors there are up to 5 calibration points available It s recommended to select them over the range of 10 100 In general 3 points should be sufficient 10 50 90 Act Flow This is the flow w...

Page 26: ...her with the compressor in the factory or on the site before a flow meter exchange 9 2 5 Downloading Settings In case S431 has to be replaced the settings of the previous installed S431 can be downloa...

Page 27: ...asuring range The standard range is the range from 0 to max 10 2 Pulse output The S431 sends out one pulse per a consumption unit by default This pulse output can be connected to an external pulse cou...

Page 28: ...ase you can set the pulse to 1 pulse per 10 consumption units or 1 pulse per 100 consumption units using by the S4C FS service app or a connected display For example with the setting of 1 pulse per 10...

Page 29: ...10 Signal outputs 10 2 1 Pulse connection diagram Using the isolated pulse switch M12 connector Pin4 and Pin5 Variant 1 Variant 2 S431 OEM 29...

Page 30: ...1 FLOAT 4 Byte 0 Medium pressure 0 010 FLOAT 4 Byte 2 Actual velocity 0 1 FLOAT 4 Byte 4 Flow 0 1 FLOAT 4 Byte 6 Consumption 1 0 UNIT32 4 Byte 8 System status N A UINT32 4 Byte 24 Remark The physical...

Page 31: ...atuses via Modbus as well The 32 bit data information is read as single bit The meanings of these bits are described as follows Bit Description Bit Description 2 Measurement over range 9 Pressure sens...

Page 32: ...the manufacturer for inspection and manufacturer 12 Disposal or waste Electronic devices are recyclable material and do not belong in the household waste The device the accessories and its packings mu...

Page 33: ...8 San Wan Road Sheung Shui N T Hong Kong Tel 49 0 7634 50488 00 Fax 49 0 7634 50488 19 Email sales suto itec com Website http www suto itec com Tel 852 2328 9782 Fax 852 2671 3863 Email sales suto ite...

Reviews: