3.2.2 Cabling and High Voltage

The general connection scheme of the CEM 4230 Detector Setup including its CEM decouple box is shown 

in 

Figure 2

. The components shown greyed out are additional possibilities for further pulse processing. 

Those devices are not part of the CEM 4230 Detector Setup.

Figure 2: Possible connection scheme of the CEM 4230 Detector Setup.

•  Connect any HV Supply to the 3 SHV feedthroughs of the CEM base flange. 
•  Connect any HV Supply to the “U-DET” SHV connector of the decoupling box.
•  In case of using a Surface Concept CEM Decoupling Box, mount the CEM Decoupling Box to the flange 

of the CEM and connect the 3 SHV cables of the decoupling box to the 3 SHV feedthroughs of the CEM 

flange. 

•  Terminate the “U-IN” SHV connector of the decoupling box with a SHV termination plug or by a defined 

reference potential. 

•  “Pulse Out” is the signal output for the CEM pulses. The corresponding connector type is BNC with a 50 

Ohm terminated signal line. 

Summary of Contents for CEM 4230

Page 1: ...CEM 4230 Detector Setup Manual CEM 4230 Detector Setup Manual...

Page 2: ...131 62716 0 fax 49 6131 62716 29 email info surface concept de web www surface concept de All rights reserved No part of this manual may be reproduced without the prior permission of Surface Concept G...

Page 3: ...lation 5 3 1 Initial Inspection 5 3 2 Installation 5 3 2 1 Mounting the CEM 5 3 2 2 Cabling and High Voltage 7 4 Operation of the CEM 9 4 1 Start Up Procedure for First Time Operation 9 4 1 1 Basic Ch...

Page 4: ...tor Follow this information to ensure a proper functioning of the detector The caution symbol marks warnings which are given to prevent an accidentally damaging of the detector or the readout system D...

Page 5: ...contact SURFACE CONCEPT immediately Please check the delivery according to the packing list see Table 1 for completeness CEM 4230 Detector Setup in transport container delivered under Argon atmospher...

Page 6: ...h the walls or other obstacles with the CEM funnel and do not pinch the cables between the flange edges After installation of the CEM assembly evacuate the chamber This is a high pressure channeltron...

Page 7: ...Connect any HV Supply to the 3 SHV feedthroughs of the CEM base flange Connect any HV Supply to the U DET SHV connector of the decoupling box In case of using a Surface Concept CEM Decoupling Box mou...

Page 8: ...ition counting and or time resolved measurements Connect any HV Supply to the 3 SHV feedthroughs of the CEM base flange Finish the complete cabling before any HV supplies and or pulse processing devic...

Page 9: ...d ground The only measurable resistance should be between R U IN U OUT of several 10 MOhm of R U IN U DET of several 10 MOhm and of R U OUT R DET of 1 MOhm All resistances towards ground should be inf...

Page 10: ...eforeperforming any changes of the cabling 4 2 Standard Operating Procedure Use this procedure later on for all operations when the detector has already been operated in vacuum before and has not been...

Page 11: ...11 CEM 4230 Detector Setup Manual 4 3 Bake Out Procedure This high pressure channeltron assembly is not intended for bake out...

Page 12: ...required Alternatively store the CEMs in argon dry air or even better in vacuum Water may lead to a reduction in the work function resulting in a higher gain During the burn in after having been stor...

Page 13: ...13 CEM 4230 Detector Setup Manual 6 List of Figures Figure 1 A CEM assembly mounted to the transport container and after removal 5 Figure 2 Possible connection scheme of the CEM 4230 Detector Setup 7...

Reviews: