Positioning  of  the CO

2

  Controller is  crucial in the  aspect  of  vibration isolation. There 

are  two  rude  vibration  sources  implemented  in  the  CO

2

  Controller  equipment.  They 

are  a  mixing  fan  and  a gas transport pump. Please  choose  such a place for the  CO

2

 

Controller  equipment,  which  provides  efficient  separation  of  it  from  the  vibration 

isolation table holding the microscope. 

 

 

Steps of Installation 

 

Set the pressure regulator on the CO

tank to a value between 1 and 2 Bar. Connect 

the outlet port of the regulator to the inlet port of the CO

2

 Controller device (using PU 

8/6 tubing). 

Fill the humidifier with water to the desired level, approximately 2/3 of the bottle. 

Connect  the  “OUT”  port  of  the  CO

2

  Controller  to  the  inlet  port  of  the  humidifier. 

Please keep the outlet and inlet ports of the humidifier bottle in mind. 

Connect  the  outlet  port  of  the  humidifier  to  the  inlet  port  of  the  incubator  (marked 

with an arrow). 

Connect the outlet port of the incubator to the “IN” port of the CO

2

 Controller device. 

Open the output valve on the pressure regulator of the CO

2

 tank. 

Plug  in  the  device  and  turn  it  on  with  the 

MAINS

  switch.  The  internal  fan  starts  up 

and  the  screen  turns  on.  The  regulator  is  ready  after  about  5  seconds.  The  CO

2

 

concentration  is  set  to  5%  by  default.  The  device  measures  CO

2

  concentration 

starting  at  2.5%,  the  display  says  0%  until  the  threshold  is  reached.  The 

concentration rises up to 6% after the first CO

2

 addition cycle. As the gases continue 

mixing the concentration drops, and as it gets below the set value, the device adjusts 

the dosing. Then the concentration stays between constant values. 

To adjust the concentration push buttons 

F

 and 

 or 

 at the same time. Release 

F

 

to save the  current setting.  This  will  be  the  new  default  the next  time  the  device is 

switched on. 

After the second cycle, the gas may be circulated through the incubator pushing the 

button 

PUMP

.  When  setting  a  lower  value  than  used  previously,  the  CO

2

 

concentration drops faster if the incubator is opened. 

 