Super Systems DPC2500 Product Manual Download Page 1

 Super Systems Inc.  Page 1 3/25/2021 

Super Systems Inc. 
Model DPC2500 –continuous dew point analyzer  

 

 

Product manual 

 

 
 
 

INTRODUCTION 

– 

 
THANKS! – 

for choosing Super Systems Inc. as your supplier and the DPC2500 as your 

continuous Dew Point Analyzer. 

 

UNPACKING 

 

Carefully unpack the Model DPC2500 Dew Point Analyzer.  If there are any signs of shipping 
damage, notify 

SSi

 

and the shipper. If a camera is available it is recommended that pictures be 

taken to show the damage to the relevant parties. 
 

NOTE: 

 

Keep this instruction book in a secure place and refer to it when there is a question about the 
analyzer

 

Summary of Contents for DPC2500

Page 1: ...500 as your continuous Dew Point Analyzer UNPACKING Carefully unpack the Model DPC2500 Dew Point Analyzer If there are any signs of shipping damage notify SSi and the shipper If a camera is available...

Page 2: ...tems Inc Page 2 3 25 2021 Table 0f contents INTRODUCTION 1 UNPACKING 1 NOTE 1 OPERATION 3 HOW IT WORKS 3 WHAT IT MEASURES 3 FACTORY CALIBRATION 5 FIELD CALIBRATION 5 FIELD CALIBRATION CONTINUED 6 SPAR...

Page 3: ...ctric ceramic that varies its electrical capacitance with changes in relative humidity The dual RH temperature sensor is mounted in a short probe which in turn is installed in a T fitting that allows...

Page 4: ...uivalent is clean and functional for generator operation you should turn the pump toggle switch OFF Maintain at least a 1 5 2 0 SCFH atmosphere flow through the analyzer Remember that the sample line...

Page 5: ...his kit consists of two bottles of saturated salt solution in which each bottle generates a precise relative humidity One bottle is 11 3 R H and the other 75 3 R H These two calibration points are alr...

Page 6: ...he bottle Step No 6 Turn the unit ON Step No 7 Press the CAL BUTTON and the SPAN BUTTON simultaneously Release both buttons See drawing Page 5 NOTE Do not be concerned if the front panel display does...

Page 7: ...0 continuous dew point analyzer Product manual SPARE PARTS Description Part No Factory Calibration 13045 Sample Pump 31401 Filter 37048 Replacement filter element 31027 Flow meter 36027 Microprocessor...

Page 8: ...n History Rev Description Date Initial Release 11 01 2000 A Added Revision History Page 07 09 2001 B SSi Address Update General Update 04 13 2005 C Corrected terminal block wiring chart 3 18 2020 D Ad...

Page 9: ...121 306 5180 http www supersystemseurope com Super Systems M xico Sistemas Superiores Integrales S de RL de CV Acceso IV No 31 Int H Parque Industrial Benito Juarez C P 76120 Queretaro Qro Phone 52 44...

Reviews: