DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc.
Page # 6 of 30
OPERATION -
The use of the DP2000 is somewhat dependent upon the application. Although
they are fundamentally alike, the operation procedures are different for sampling
heat treating furnaces and endothermic generators.
Flip the POWER switch to the on position. A green POWER ON light should
illuminate indicating that the unit is turned on. The LED display should also
illuminate showing “SSi” (Figure 1) followed by a dew point reading given in
degrees Fahrenheit (Figure 2).
Figure 1 Figure 2
Connect a sampling tube/hose to the barb fitting located on the side of the
DP2000 (Figure 3). Make sure that the hose or tubing fits over the fitting properly
and does not allow air to leak into the sampling line. A tubing and filter assembly
is supplied with the DP2000 (SSI P/N A20315 – Figure 5), which is designed to fit
the hose barb properly. Also, make sure that the sample gas is flowing through
the filter in the right direction. There is an arrow on the filter to indicate which
direction the gas should flow (Figure 4).
Figure 3 Figure 4