PROCEDURE FOR CHECKING SET VALUES

Set to Set Value 3 (P3) set mode

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

Displayed 
alternately

In the Set Value 2 (P2) set mode, press 

key.

• 

and the automatically set Set Value

3 (P3) are displayed alternately.

• 

In case
Set Value 1 (P1) = 

50.0kPa

Set Value 2 (P2) = 

100.0kPa

then

Set Value 3 (P3) =

50.0

(

100.0) 

75.0kPa

2

Digits smaller than the displayed digits
are discarded.

MODE

In case Set Value 3 (P3) is to be changed

The automatically set Set Value 3 (P3) can be manually changed to
a value between Set Value 1 (P1) and Set Value 2 (P2). However
Set Value 3 (P3) cannot be set to a value on the vacuum pressure
side for a positive pressure type sensor or to the positive pressure
side for a vacuum pressure type sensor.

Set to sensing mode

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

Press       key.

• 

The sensor returns to the sensing mode
after Set Value 1 (P1), Set Value 2 (P2)
and Set Value 3 (P3) have been set.

• 

Since the values which have been set are
stored in an EEPROM, they are not erased
even if the power supply is switched off.

MODE

The conditions which have been set in the initial setting and the
pressure settings can be checked by the following procedure.

Please note that if any key, except      key, is pressed in any
setting mode, the set conditions shall get changed.

MODE

Procedure to check initial conditions

Procedure to check set values

Set to sensing mode

Set to sensing mode

Press       key
while pressing 

key.

Press       key.

Sensor enters initial setting mode

Sensor enters Set Value 1 (P1) set mode

• 

Initial conditions which have
been set are displayed and can
be checked.

• 

Set Value 1 (P1) which has
been set is displayed and can
be checked.

Sensor returns to sensing mode

Sensor enters Set Value 2 (P2) set mode

• 

Set Value 2 (P2) which has
been set is displayed and can
be checked.

Sensor enters Set Value 3 (P3) set mode

• 

Set Value 3 (P3) which has
been set is displayed and can
be checked.

Sensor returns to sensing mode

MODE

Press       key.

In case of auto-
matic sensitivity
setting mode only

MODE

MODE

Press       key.

MODE

Press        key.

MODE

Press       key.

MODE

13

CONVERSION OF PRESSURE UNITS

In the DP2 series, the conversion to different units is automatically done on changing the setting of the pressure unit. However,
this conversion can also be obtained by multiplying the values by the coefficients given in the following table.

For example, if 2kPa is to be expressed in kgf/cm

2

, since 1kPa=1.01972

10

2

kgf/cm

2

, 2kPa becomes 2

1.01972

10

2

c

0.020kgf/cm

2

Conversion procedure

Conversion table for pressure units

kPa

MPa

kgf/cm

2

bar

psi

mmHg

inHg

atm

(Torr)

1kPa

1

1

10

3

1.01972

10

2

1

10

2

1.45038

10

1

7.50062

0.2953

9.86923

10

3

1MPa

1

10

3

1

1.01972

10

1

10

1.45038

10

2

7.50062

10

3

0.2953

10

3

9.86923

1kgf/cm

2

9.80665

10 9.80665

10

2

1

9.80665

10

1

1.42234

10 7.35559

10

2

2.8959

10 9.67841

10

1

1bar

1

10

2

1

10

1

1.01972

1

1.45038

10 7.50062

10

2

2.953

10

9.86923

10

1

1psi

6.89473

6.89473

10

3

7.03065

10

2

6.89473

10

2

1

5.17147

10

2.036

6.80457

10

2

1mmHg

1.33322

10

1

1.33322

10

4

1.35951

10

3

1.33322

10

3

1.93368

10

2

1

3.9370

10

2

1.31579

10

3

(1Torr)

1inHg

3.3864

3.3864

10

3

3.4531

10

2

3.3864

10

2

0.4912

2.5400

10

1

3.342

10

2

1atm

1.01325

10

2

1.01325

10

1

1.03323

1.01325

1.46960

10 7.60000

10

2

2.9921

10

1

14

PEAK HOLD & BOTTOM HOLD FUNCTIONS

15

KEY-PROTECT FUNCTION

Peak hold and bottom hold functions enable the display of the
peak value (maximum pressure value in case of the positive
pressure type sensor and maximum vacuum pressure value
in case of the vacuum pressure type sensor) and the bottom
value (minimum pressure value in case of the positive
pressure type sensor and minimum vacuum pressure value in
case of the vacuum pressure type sensor) of the varying
measured pressure.

These functions are convenient for finding the pressure
variation range or for determining the reference for pressure
settings.

Key-protect is a function which prevents any unintentional
change in the conditions which have been entered in each
setting mode by making the sensor not to respond to the key
operations.

16

LABEL FOR CHANGE IN PRESSURE UNIT

When a pressure unit other than ‘kPa’ or ‘MPa’ has been
selected in the initial setting mode, the label (supplied as
accessory) which corresponds to the selected unit should be
stuck at the position shown in the figure below.

Please note that the peak value and the bottom value data
is erased when it is no longer displayed.

The response time of the comparative outputs becomes
slower during the peak hold and bottom hold display.

Since the key-protect information is stored in an EEPROM,
it is not erased even if the power supply is switched off.

Please take care to remember if the key-protect function
has been set.

Peak hold display

Initiating peak hold display

Setting of key-protect

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

Displayed 
alternately

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

In the sensing mode, keep     key pressed
until         is displayed. (4 sec. approx.)
When the finger is released after      is
displayed, the peak value and       are
displayed alternately.

• 

If the applied pressure exceeds the
displayable pressure range, error
message (       or       ) and        are
displayed alternately. In this case, bring
back the applied pressure to within the
rated pressure range.

• 

The figure on the left shows the display of
a vacuum type sensor when the pressure
unit has been set to ‘kPa’.

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

In the sensing mode, press     key
continuously for about 3 sec. and release
it immediately when          is displayed.

• 

Key-protect is set and the sensor returns
to the sensing mode.

MODE

Ending peak hold display

Press       key.

[

• 

Sensor returns to sensing mode.]

Bottom hold display

Initiating bottom hold display

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

Displayed 
alternately

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

In the sensing mode, keep     key pressed
until         is displayed. (4 sec. approx.)
When the finger is released after       is
displayed, the bottom value and        are
displayed alternately.

• 

If the applied pressure exceeds the
displayable pressure range, error
message         or        and       are
displayed alternately. In this case, bring
back the applied pressure to within the
rated pressure range.

• 

The figure on the left shows the display of
a vacuum type sensor when the pressure
unit has been set to ‘kPa’.

Ending bottom hold display

Press       key.

[

• 

Sensor returns to sensing mode.]

When the keys are to be operated, make sure that key-
protect is released.

Release of key-protect

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

In the sensing mode, press     key
continuously for about 3 sec. and release
it immediately when           is displayed.

• 

Key-protect is released and the sensor
returns to the sensing mode.

MODE

Pressure unit label (accessory)

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

Comparative 
Output 1

Comparative 
Output 2

ON

OFF

ON

OFF

0 Set Value 1 (P1)

Set Value 2 (P2)

Set Value 3 (P3) High pressure 

(Positive pressure type)
High vacuum 
(Vacuum pressure type)

Hysteresis

Changeable 
range

Hysteresis: 1 digit

2 digits when 
psi unit is used

(

)

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

DP2-20

–101.3kPa

OUT2

OUT1

MODE

0-ADJ

Displayed 
alternately

Enter using       key and       key.

In case of a positive pressure type
sensor, if     key is pressed once, the set
value changes towards the high pressure
side by 1 digit and if     key is pressed
once, the set value changes towards the
low pressure side by 1 digit.
In case of a vacuum pressure type
sensor, if      key is pressed once the set
value changes towards the high vacuum
side by 1 digit and if     key is pressed
once the set value changes towards the
low vacuum side by 1 digit.

If     key or     key is pressed conti-
nuously, the set value changes quickly.

• 

If the set pressure range is exceeded,
either       (upper limit exceeded) or 

(lower limit exceeded) is displayed.

(

)