MANAGING READINGS
MANAGING READINGS
MANAGING READINGS
MANAGING READINGS
Recalling the Last Set of Readings
Recalling the Last Set of Readings
Recalling the Last Set of Readings
Recalling the Last Set of Readings
To redisplay the last set of readings, depress the automatic button
on the BP module for
more than two seconds until the last reading is displayed. If your
SunTech 247
has
temperature and/or pulse oximetry modules, the last set of readings includes these readings as
well. If the last attempted reading resulted in an error and/or warning, then this will be
displayed. The device will display dashes if no reading is in memory, a reading was aborted
or, the previous BP was a Sphygmode measurement.
For the pulse oximetry module: In the event of the 10-minute measurement timeout, the
module will terminate the measurement and status code “01” will display on the main
module. The last valid reading recorded at the end of the ten-minute period will be the
recalled reading.
Clear
Clear
Clear
Clearing
ing
ing
ing the Last
the Last
the Last
the Last Set of Readings
Set of Readings
Set of Readings
Set of Readings
To clear the values from the last automatic BP measurement and the accessories, press and
hold the automatic button
more than 5 seconds. Previous values will be displayed
momentarily. Then the display blanks. On redisplay you will see dashes for all the values
that have been cleared.
Your
SunTech 247
displays the most recent set of readings for two minutes.
If patient privacy is a concern, you can clear these readings from the display
before collecting vital signs from another patient.