SUNSTONE Orion M Series User Manual Download Page 10

m Series 

User Manual

10

INSTALL THE TUNGSTEN ELECTRODE ONTO THE WELDING STYLUS

Follow these steps to properly install the tungsten electrode . 

• 

Remove the stylus hull by pulling it away from the stylus shaft .

• 

Loosen the collet cap by twisting it counter-clockwise . 

• 

The welder comes with 2 electrode collets . One that fits 0 .5mm 
electrodes and one that fits 1 .0mm electrodes . The electrode stylus will 
be shipped with the 1 .0mm electrode collet installed . 

• 

Insert a 1 .0mm electrode into the collet .  Helpful Tip: There is a groove cut 
around the stylus hull that will help measure the electrode length . Place 
the end of the stylus hull up against the collet cap, then make sure the 
electrode tip falls between groove . 

• 

There should be between 0 .6 - 0 .7in (1 .5 – 2cm) of the electrode 
protruding from the stylus shaft . This will allow the electrode enough 
room to stick out from the stylus once the stylus hull is placed back on 
the stylus .)

• 

Lock the electrode into place by hand tightening the collet cap in a 
clockwise direction .

• 

Replace the stylus hull by pushing it in until you feel it snap back into 
place (the electrode should stick out between 1/8 – 1/4in (3 .75 – 
6 .75mm) after the stylus hull is snapped back into place) .

WORK PIECE TO ELECTRODE PRESSURE

Touch the work piece to the electrode with very light pressure . Too much pressure will cause the work piece to stick to 
the electrode and in turn cause the electrode to be contaminated (work piece material on the electrode) . We recommend 
cleaning or changing the electrode any time it gets stuck to the work piece

WHEN TO SHARPEN THE ELECTRODE

It is recommended that the user pay close attention to the electrode condition .  An 
electrode that appears to be dark colored or covered with material from previous 
welds can lead to inconsistent welding and poor igniting of the weld .  When this occurs, 
simply sharpen the electrode with the included diamond disk . The diamond disk can 
be attached to a flex shaft or Dremel tool . Follow these steps for sharpening the 
electrode .

• 

Completely remove the electrode from the stylus .

• 

Pinch the electrode between the thumb and index and/or middle finger with the 
shaft going perpendicular to the fingers .

• 

Power on the Dremel or flex shaft then hold it with the opposite hand .  *Place the 
electrode tip in the diamond disc in such a way that the grit of the disk is moving 
parallel with the electrode shaft and moving towards the tip .

• 

This will affect the quality of the weld if not done as explained above .

• 

Set the electrode on the diamond disk at a 10-degree angle and begin to spin 
the electrode with the thumb and middle finger . *A helpful way to get a sharp 

Summary of Contents for Orion M Series

Page 1: ...Orion m Series Welding System User Manual...

Page 2: ......

Page 3: ...15 Chapter 3 Make A Weld p 17 Make an Arc Weld p 17 Orion mPulse Pulse Arc Welder Welder Type Pulse Arc Weld Modes 1 Pre Programmed Metal Settings 0 Customizable Save Settings 0 Languages None Displa...

Page 4: ...m Series User Manual 4...

Page 5: ...supply that can accept both 120 and 240VAC The darkening lens or microscope will automatically darken when the weld takes place This allows the user to have a view of the work piece and protection fro...

Page 6: ...p to table Determine the height of the tabletop If it is thicker than 3 1 2 9cm follow mounting options 2 or 3 Position the microscope arm clamp under the tabletop Turn the clamp knob clockwise until...

Page 7: ...round clamp to secure the microscope to the arm Install the rubber eyepiece covers onto the microscope lenses Insert the welding stylus into the stylus holder under the microscope then tighten the st...

Page 8: ...he stylus holder by tightening the screw on the bottom of the stylus holder ADJUST THE MICROSCOPE FOCUS Twist the knob on the microscope forward and backward to raise and lower the head This will allo...

Page 9: ...nts It is helpful to have your hands resting and to only use your fingers to move the work piece up to the electrode Place the work piece surface perpendicular to the point of the electrode As discuss...

Page 10: ...CE TO ELECTRODE PRESSURE Touch the work piece to the electrode with very light pressure Too much pressure will cause the work piece to stick to the electrode and in turn cause the electrode to be cont...

Page 11: ...electrode Once the electrode has a flat blunt tip turn the Dremel off and insert the electrode back into the stylus as explained above Shielding Gas Setup During the pulse arc welding process high tem...

Page 12: ...nect one end of the gas tubing to the gas regulator Insert the other end of the gas tubing into the gas port on the back of the power supply It will stop when it is fully connected Tug gently on the t...

Page 13: ...13 OrionWelders com...

Page 14: ...m Series User Manual 14...

Page 15: ...SE BUTTONS Play In order to weld the play pause button must show the green triangle play symbol Pause If the welder is not going to be in use for a period of time users can push the play pause button...

Page 16: ...m Series User Manual 16...

Page 17: ...verify the gas pressure and microscope connection 2 Lightly touch the electrode in the box on the provided weld plate Maintain contact keep hands steady by resting them on the table 3 Use the provide...

Page 18: ...nergy Notes 14k Gold Sharp 30 ws Butt weld with 90 degree electrode angle 24k Gold Sharp 30 ws Silver Blunt 30 ws Platinum Sharp 30 ws Stainless Steel Sharp 30 ws Palladium Sharp 30 ws Titanium Sharp...

Page 19: ...tinum Sharp 20 ws Stainless Steel Sharp 15 ws Palladium Sharp 15 ws Titanium Sharp 20 ws Brass Sharp 17 5 ws Retip Prong 26 AWG wire Metal Tip Shape Energy Notes 14k Gold Sharp 10 ws Butt weld with 90...

Page 20: ......

Reviews: