10
• Questions? Call or Text +1-801-658-0015
LZR LASER WELDER USERS GUIDE
• Wait about 60 seconds, then turn the key and press start. If an error message appears saying: “ERROR
02 TEST FLUX H20” then follow these instructions:
• Turn the key to the off position.
• Turn the welder off using the power switch on the rear panel.
• Turn the welder on using the power switch on the rear panel.
• Push the internal joystick down and hold down for 5 to 6 seconds or until you hear the welder’s
fans and the pump turn on. The air will be forced out of the system. You should not see the error
message. If the error message persists, call Sunstone’s customer service team at +1-801-658-
0015 for assistance.
• Turn the key and press the Start Button which will appear on the display screen.
• If you do not see the “ERROR 02 TEST FLUX H20” error, your welder is ready to use. The final step is to
plug the Upper Connector with the Red Plug, as seen in Figure 9.4.
Powering the Welder for the First Time
• Ensure that the Emergency Stop Button is disengaged
and not pushed in. See Figure 10.1.
• Turn on the welder by flipping the switch on the back of
the welder.
• Insert the key and turn clockwise so the key is vertical.
See Figure 10.2.
• A Start Button will appear in the touchscreen interface.
Press the Start Button to adjust weld parameters and to
weld. See Figure 10.3.
• When you have pressed “START” on the display the
default set values will appear and the inscription “VERI-
FICATION VALUES LASER ACTIVE”.
NOTE: Do not proceed with the installation if the display does
not show as described above. If you encounter any problems or
errors, consult chapter 5 of this document, or contact Sunstone
Engineering.
At this point you will be able to set welding parameters, load
saved settings, if desired, and begin welding.
To turn the welder off, turn the key to the Locked position, wait
for the touchscreen to display the message “LASER LOCKED,”
then flip the rear power switch to the off position.
ATTENTION: When the welder is not being used, ensure that the
key is removed from the welder.
Figure 10.1. Make sure the
Emergency Stop Button is
disengaged.
Figure 10.2. Turn the key
clockwise to access the
touchscreen controls.
Figure 10.3. Touch the Start Button to adjust weld
parameters and start welding.