![Sunrise Medical Oxford MAJOR 190 User Instruction Manual Download Page 22](http://html1.mh-extra.com/html/sunrise-medical/oxford-major-190/oxford-major-190_user-instruction-manual_1399493022.webp)
O x f o r d R a n g e o f M o b i l e H o i s t s
•
EMERGENCY DOWN:
Check emergency down function.
•
HYDRAULIC UNIT:
The hydraulic unit should require no maintenance other than
checking for correct operation and leakage of hydraulic fluid.
•
BATTERIES (Electric only):
The batteries are located in the Power Pack and should
not require maintenance other than the regular charging as detailed in the charging
instructions.
•
CLEANING:
Clean with ordinary soap and water and/or any hard surface
disinfectant. Harsh chemical cleaners or abrasives should be avoided as these may
damage the surface finish of the lift. Avoid wetting any of the electrical parts.
LOAD TEST:
The load test should be carried out in accordance with the manufacturers
test procedures. It is strongly recommended the testing is carried out by an authorised
service dealer.
CERTIFICATION:
An authorised service dealer will issue a test certificate after
satisfactory completion of the load test. This certificate will be valid for six months.