1.  INTENDED USE

 

 

The battery and charger are intended for use with the LED Headlight System which is 
intended to deliver illumination from a high intensity LED for surgical site illumination. 

2.  WARNINGS

 

  Upon initial receipt and before each use inspect each Battery and Charger for damage.  

Do not use any Battery if it appears to be damaged.  Do not use a Battery pack that gives 
off an odor, generates heat, becomes discolored, or leaks.  Do not use the Charger if the 
cord and/or plug appear to be damaged, such as cuts, bent pins or contact, and/or cracks. 

  Do not expose the Battery or Charger to heat, fire, or mechanical shock.  Do not heat 

the Battery above 80°C (176° F).   

  Exposure to elevated temperatures may reduce Battery capacity and performance.  Do 

not expose the Battery or Charger to heat sources or sunlight.  Warm environments can 
cause the calibration to fail. Place the Charger in a cool spot, away from external heat 
sources. 

  Do not cover the fan exhaust or obstruct the airflow. This will cause overheating. 

  Do not use this equipment in the presence of a mixture containing a flammable 

anesthetic and/or air or oxygen or nitrous oxide. 

  Do not sterilize, immerse, or expose the Battery, Charger, or Power Supply to water, 

solvents, lubricants, or other chemicals to clean the Charger unless otherwise directed.  
Do not allow water to collect in the modules or on top of the Charger.  Do not allow water 
to enter the power connection on the back of the Charger.  Keep clean and dry.  

  Do not short-circuit, crush, open, shred, incinerate, or dismantle the Battery, Charger, 

or Power Supply case. Do not store cells or Batteries haphazardly in a box or drawer 
where they may short-circuit each other or be short-circuited by other metal objects.  
Damage increases the risk of chemical burns from electrolyte leakage, fire or explosion.   
The Charger does not have serviceable parts. Please contact Cuda Surgical for 
replacement instructions.   

  Do not modify or change the Charger Power Cord.  The Power Cord should be plugged 

into a hospital grade outlet. 

  Use only the Battery, Charger, and Charger accessories in the application for which 

they are intended. 

  Do not operate the Charger at a different voltage than what is listed on the unit and in 

the Instructions For Use. 
 

3.  BATTERY HANDLING RECOMMENDATIONS

 

 

  Do not remove a Battery from its original packaging until required for use. 

  Observe  the  plus  (+)  and  minus  (

–)  marks  on  the  Battery  and  ensure  correct 

orientation. 

  After  extended  periods  of  storage,  it  may  be  necessary  to  charge  and  discharge  the 

Battery several times to obtain maximum performance. 

  Dispose Battery and Charger per Disposal/Recycle section of this Instructions For Use. 

  Rechargeable  Batteries  give  the  best  performance  when  they  are  operated  at  normal 

room temperature, 20 °C ± 5 °C (68°F ±10°F). 

  Retain the original product literature for future reference. 

Summary of Contents for LIT-210

Page 1: ...Page 1 of 8 6 Bay Battery Charger Operation Manual Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville FL 32216 USA Customer Service 904 737 7611 Toll Free 877 677 2832 EC REP RMS UK Ltd 28 Tri...

Page 2: ...OF CONTENTS 1 INTENDED USE 2 GENERAL WARNINGS 3 BATTERY HANDLING RECOMMENDATIONS 4 BATTERY CHARGING INSTRUCTIONS 5 LED INDICATION 6 BATTERY RECALIBRATION INSTRUCTIONS 7 CLEANING INSTRUCTIONS 8 STORAGE...

Page 3: ...unless otherwise directed Do not allow water to collect in the modules or on top of the Charger Do not allow water to enter the power connection on the back of the Charger Keep clean and dry Do not s...

Page 4: ...Charge Figure 2 Battery Charger 1 Place the Charger on a flat level surface away from sources of heat and moisture 2 Connect the Power Supply to the mains AC supply using the cord provided 3 Place the...

Page 5: ...the Charger may become warm Calibration cycle will perform best if the Battery has a full charge in the beginning Battery can either be calibrated and charged or just charged Calibration takes longer...

Page 6: ...ION Recommended Storage and Transportation Conditions 0 F 18 C to 113 F 45 C up to 80 RH CAUTION For optimal Battery life the Battery packs should be stored in an environment with low humidity free fr...

Page 7: ...Charger may have a faulty Power Cord Replace the Power Cord Charger LED indicator not working Ensure Charger Power Cord is plugged in to the AC power outlet Ensure Battery is properly seated in the C...

Page 8: ...how to properly dispose of this product please contact your local Zimmer Representative Contains a Li ion Battery that must be recycled Contains a Li ion Battery that must be recycled This product con...