Sunnex PF Series MRI Service & Installation Manual
26
Recommended Separation Distances Between Portable and
Mobile RF Communications Equipment and the PF MRI Light (PF
28016-M) Equipment and Systems that are NOT Life-Supporting
The PF MRI Light (PF 28016-M)
is intended for use in the
electromagnetic environment in which radiated disturbances are
controlled. The customer or user of the PF MRI Light (PF 28016-M)
can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF Communications
Equipment and the PF MRI Light (PF 28016-M) as recommended
below, according to the maximum output power of the
communications equipment.
Recommended Separation Distances for the PF
MRI Light (PF 28016-M)
Maximum
Output Power
(Watts)
150 kHz to 80
MHz
d =
1.1667
P
80 MHz to
800MHz
d =
1.1667
P
800 MHz to 2.5
GHz
d =
2.3333
P
0.01 0.11667 0.11667 0.23333
0.1
0.36894 0.36894 0.73785
1
1.1667 1.1667 2.3333
10
3.6894 3.6894 7.3785
100 11.667 11.667 23.333