User Guide

Instruction for use, continued

6.  To add prescribed medication, do not unscrew medication chamber. 

Remove the inspiratory valve and instill medication through the top (D).
•  Any medication indicated for use with small-volume jet nebulizer  

can be used with the NebuTech.

•  Maximum medication volume is 5 mL.
•  Do not instill medication through the side elbow because medication  

can be caught in the drool catcher.

•  Replace inspiratory valve after adding medication (E).

7.  To start treatment set flow rate at 4 LPM to 8 LPM  

(25 PSI to 52 PSI)

•  For optimal performance use a flow rate of 8 LPM.
•  3.0 mL of medication will nebulize approximately 4 minutes @ 8 LPM.

8.  After treatment, shake out excess fluid.
9.  Follow hospital’s policy for storage and cleaning  

of nebulizers between treatments.

Treatment Tips

1.  If the nebulizer does produce a heavy aerosol, tap the sides of the 

medication chamber to ensure the green cone is seated properly. Ensure 
that the medication cup is screwed on tightly to the reservoir tower.

2.  Instruct patient to keep their lips sealed around mouthpiece during 

inspiration and exhalation.

3.  Ensure nebulizer is in a position upright or greater than  

a 45 degree angle.

4.  After each treatment, shake out any residual medication.
5.  If the nebulizer has been in use for the several days, the capillary tube  

of the nebulizer may be partially occluded. Replace nebulizer.

6.  Ensure the exhalation valve on mouthpiece is upright to prevent moisture 

dripping on patient.

NebuTech® 8960 Disposable Nebulizer Cont.

Disposable | Single Patient Use

D

E

Summary of Contents for NebuTech 8960

Page 1: ...or single patient use and is not intended to be reprocessed or autoclaved Follow your facilities policies for cleaning and changing nebulizers If the nebulizer is used for more than one day consider writing the date on the equipment bag or nebulizer tower with a permanent marker Instructions for use 1 Wash your hands Use an aseptic technique to assemble the nebulizer 2 Remove the NebuTech Nebulize...

Page 2: ...er treatment shake out excess fluid 9 Follow hospital s policy for storage and cleaning of nebulizers between treatments Treatment Tips 1 If the nebulizer does produce a heavy aerosol tap the sides of the medication chamber to ensure the green cone is seated properly Ensure that the medication cup is screwed on tightly to the reservoir tower 2 Instruct patient to keep their lips sealed around mout...

Page 3: ...cluded in package to connect the mask to the nebulizer 5 Set the flow rate from 5 LPM to 8 LPM 25 PSI to 52 PSI to begin the treatment 6 Apply the mask to the patient Alternative Mask Placements If the patient is unable to sit upright or is in a semi fowlers position attach the mask to the nebulizer s side port Setup A 1 Connect mask to the side port Setup B 1 Additional equipment is required Gath...

Page 4: ...every 7 days or after 21 cleaning cycles whichever is first Replace the nebulizer sooner if the it becomes damaged contaminated or is not functioning properly Cleaning Method 1 After each use thoroughly rinse the nebulizer with sterilized water DO NOT submerge the supply tubing in water 2 Disassemble all nebulizer components do not remove any of the one way valves 3 Rinse with sterilized water sha...