SPECIFICATIONS
EXPLANATION OF SYMBOLS
Model
HL858A2-LT
Internal power supply
DC 6V (4 x 1.5V AAA alkaline batteries)
DC 5V 1A (AC/DC Adapter)
Measurement method
Oscillographic mode (inflation)
Measurement range of Cuff
Pressure 0 - 40kPa (0 - 300mmHg)
Measurement range of Determination
Pressure 5 - 39kPa (40 - 280mmHg)
Pulse value
40 - 199 beats/minute
Accuracy
Pressure ±0.4kPa (±3mmHg), Pulse value ±5%
Memory
2 x 120
Display mode
Liquid Crystal Display
Operating conditions
Temperature 41˚F - 104˚F (5˚C - 40˚C),
Relative humidity 15% - 93%, Pressure 700 - 1060hPa
Storage & transportation conditions
Temperature -13˚F - 158˚F (-25˚C - 70˚C), Relative humidity ≤93%
Unit Dimensions
5.5” x 4.3” x 2.2” (L x W x H) 14cm x 11cm x 5.6cm (L x W x H)
Unit Weight
9.74oz ± 0.18oz (276g ± 5g) (Cuff and batteries excluded)
Cuff Size
UC-01: Standard size cuff 9 - 17” (23 - 43cm)
ELC-01: Extra large size cuff 17 - 22” (43 - 56cm)
Symbol
Explanation
Health & Life Information
Refer to instruction
manual / booklet
TYPE BF Applied Part
To avoid inaccurate results caused
by electromagnetic interference
Warning: Portable RF communications equipment (in-
cluding peripherals such as antenna cables and external
antennas) should be used no closer than 30cm (12
inches) to any part of the device, Otherwise, degradation
of the performance of this equipment could result.
Waste of electrical and electronic
equipment (WEEE)
Discard the used product to the recycling collection point
according to local regulations.
Serial number
YYMMXXXXXX
Ingress Protection Rating
First characteristic numeral- Degree of protection against
access to hazardous parts and against solid foreign objects
N1=2 (Protected against solid foreign objects of 12.5mm
Ø and greater) Second characteristic numeral - Degree
of protection against ingress of water N2=2 (Protected
against vertically falling water drops when ENCLOSURE
tilted up to 15°)
Non-ionizing electromagnetic
radiation
NOTE:
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there
is no guarantee that interference will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception, which can be determined by
turning the equipment o and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and the receiver.
• Connect the equipment into an outlet on a circuit dierent from that to which the receiver
is connected.
• Consult the dealer or an experienced radio/TV technician for help.
CAUTION:
To assure continued FCC compliance: Any changes or modications not expressly approved by
the grantee of this device could void the user’s authority to operate the equipment.
RF EXPOSURE WARNING
1. The equipment complies with FCC RF exposure limits set forth for an uncontrolled
environment.
2. The equipment must not be co-located or operation in conjunction with any other antenna
or transmitter.
FCC LABEL COMPLIANCE STATEMENT:
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause undesired operation.
HL858A2-LT ESSENTIAL PERFORMANCE PER IEC 80601-2-30 ADDITIONAL
ESSENTIAL PERFORMANCE REQUIREMENTS:
• 201.12.1.102 Limits of the error of the manometer from environmental conditions. Over
the temperature range of 41°F - 104°F (5°C - 40°C) and the relative humidity range of
15% - 93%(non-condensing), the maximum error for the measurement of the CUFF
pressure at any point of the NOMINAL measurement range shall be less than or equal to
±3mmHg (±0.4kPa) or 2% of the reading, whichever is greater.
• 201.12.1.107 Reproducibility of the blood pressure determination.
The laboratory Reproducibility of the BLOOD PRESSURE DETERMINATION of the AUTOMATED
SPHYGMOMANOMETER shall be less than 3mmHg (0.4kPa).