SYMBOLS / SYMPTOMS

CONDITIONS / CAUSES

INDICATION / CORRECTION

Unit does not turn

on when 

START / STOP

button is pushed.

Worn-out batteries.

Replace them with 4 new AAA 

(LR03) alkaline batteries.

Battery polarities have been positioned 

incorrectly.

Re-insert the batteries in the correct 

positions.

Measuring Error Symbol

appears when blood

pressure value displayed

is excessively low or high.

Cuff has been placed incorrectly.

Wrap the cu properly so that it is 

positioned correctly.

Measuring 

Error Symbol

Air circuit abnormality. Cuff tube may 

not be plugged into monitor correctly.

Check cuff connection.

Measure again.

Measuring

Error Symbol

Ination pressure exceeding 300mmHg. Switch the unit o, then measure

again.

Measuring 

Error Symbol

Can’t determine blood pressure

measurement data.

Wrap the cuff properly and keep

steady. Measure again.

Cuff is worn improperly, or the shape of 

the upper arm is unusual (for example, 

the circumference of the upper arm 

diers largely from the circumference of 

the forearm), excessive gap might exist 

between the arm cuff and the arm.

Wrap the cuff snugly so that it is

positioned correctly. If you have any 

question about the cuff wearing 

and/or measurement result, 

please consult your healthcare 

professional.

Excessive Body Motion

Detector

Notice:

 The measured blood 

pressure reading may not 

be accurate if the icon is 

displayed.

Body movement during measurement, 

especially, the movement on the arm 

the blood pressure monitor is worn 

on. e.g. Talking, moving or shaking 

of the arm with the cuff on while 

measurement.

Measure again. Keep arm steady 

during measurement.

Cu is worn improperly, or the shape of 

the upper arm is unusual (for example, 

the circumference of the upper arm

diers largely from the circumference of 

the forearm), excessive gap might be 

exist between the arm cuff and the arm.

Wrap the cuff properly and keep 

steady. Measure again. If you have 

any question about the cuff wearing 

and/or measurement result, 

please consult your healthcare 

professional.

Cuff Fitness 

detection Symbol

The cuff was wrapped incorrectly (for 

example too loosely or too tightly).

Please reference “applying the Cuff“ 

section to wrap the cuff correctly

Note:

 If "EP" appears on the display, just return the device to your local distributor or importer.

RECALLING DATA

 

1.  Press 

+

 button to select User 1 or 2. 

2.  Press 

M

 button to enter Memory Mode. 

•  If there is no data stored, nothing (except month, date, and time) 

will appear on the display.  

•  If there is data stored, the first reading will show the average of the 

last three (3) records in memory. 

3.  Press 

M

 to read the preceding measurements in sequence. 

4.  When finished, press 

START / STOP

 (the device will enter Standby Mode).

ERASING DATA

1.  Press 

+

 button to select User 1 or 2. 

2.  Press 

M

 button to enter Memory Mode. 

3.  Press and hold 

M

 and 

+

 buttons at the same time. All data for the 

selected user will be erased automatically.

Note:

 Once deleted, your data 

CAN NOT

 be restored.

CLEANING AND DISINFECTION

•  Clean the device with only a soft, damp cloth. Wipe dry. 
•  Never immerse the device or cuff in water or other liquid. 
•  Do not use detergent or any strong chemicals to clean the device. 
•  Make sure the cuff is clean and completely dry before use.  
•  Inspect the device, LCD display and cuff(s) before and after each use. Discard immediately if 

signs of blood, soil or stains exist. 

•  For at-home use by one user: Clean and disinfect the device and cuff(s) at least once a 

month with a commercially available 75% alcohol wipe for a minimum of 30 seconds. 

•  For at-home use by multiple users, clean and disinfect between each user and at least once 

weekly.  

STORAGE 

•  Remove the batteries if storing or not using for an extended period. 
•  Make sure the device is clean and dry before storage. 
•  Always store and transport the device in its protective case.  
•  Avoid dropping, intense shaking or other impact. 
•  Store the device away from direct sunlight, dust, humidity, and temperature extremes.

LIMITED WARRANTY

This product has a limited warranty of 3 year(s) on the blood pressure monitor, 2 year warranty 
on the cuff and accessories (except batteries) from the original date of purchase against 
workmanship and defects in material. If under normal use, your product fails to operate, please 
contact our customer service department at [email protected]. A refund or replacement will 
be provided to you with proof of purchase. Star Elite Inc. may deny claims of damage caused by 
misuse or modifications of this product. 

TROUBLE SHOOTING

Summary of Contents for 16994

Page 1: ...1 Upper Arm Blood Pressure Monitor with Voice Broadcast Technology Instruction Manual Model HL858A2 LT Item 16994 Manual Version 001 Issue Date XX ...

Page 2: ...curate idea of your blood pressure MEASUREMENT METHOD HL858A2 LTAutomatic Upper Arm Blood Pressure Monitor measures blood pressure and heart rate by oscillometric method meaning the fluctuations in pressure are measured Once the cuff is wrapped around your upper arm just turn on the monitor and inflation automatically starts The inflation of the cuff creates pressure around the arteries inside upp...

Page 3: ...osis diabetes liver disease kidney disease severe hypertension peripheral circulation please consult your healthcare professional before using the device Blood pressure measurements taken with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method and are within the accuracy limits prescribed by the Standard of EN 1060 4 If the cuff is wor...

Page 4: ...perly BP Category Indicator The arrowhead points out the specific BP Category that your measurement reading fits in Average of Last 3 Measurements ThissymbolappearswhenLCDdisplaysaverage value of last3readings IRREGULAR HEARTBEAT DETECTOR The symbol will appear on screen indicating a certain heartbeat irregularity was detected during measurement The heartbeat rhythm that is more than or less than ...

Page 5: ... Insert 4 new AAA 1 5 V LR03 alkaline batteries into the compartment as shown Make sure the polazrities and align with the symbols inside the compartment 3 Replace the batteries when The low battery icon appears The display is dim or does not turn on Do not mix new and used batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries Such action may shorten the battery life...

Page 6: ...uff inanyway MEMORY FUNCTION 1 Themonitorfeaturesmemorystoragewithdateandtimestampsfortwo 2 users saving 120recordsperuser 240total 2 Aftereachmeasurement thesystolicanddiastolicpressure heartrateandBPCategory Indicator IrregularHeartbeatDetectorandExcessivebodymotiondetector if any withthe dateandtimewillbeautomaticallystored Eachnewreadingwillreplacetheoldest TAKING THE MEASUREMENT 1 With the cu...

Page 7: ...ol The cuff was wrapped incorrectly for example too loosely or too tightly Please reference applying the Cuff section to wrap the cuff correctly Note If EP appears on the display just return the device to your local distributor or importer RECALLING DATA 1 Press buttontoselectUser1or2 2 PressM buttontoenterMemoryMode If thereisnodatastored nothing exceptmonth date andtime willappearonthedisplay If...

Page 8: ...hthelimitsforaClassBdigital device pursuanttoPart15of theFCCRules Theselimitsaredesignedtoprovidereasonable protectionagainstharmfulinterferenceinaresidentialinstallation Thisequipmentgenerates usesandcanradiateradiofrequencyenergyand if notinstalledandusedinaccordancewith theinstructions maycauseharmfulinterferencetoradiocommunications However there isnoguaranteethatinterferencewillnotoccurinapar...

Page 9: ...15kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges AmbientTemperature 15 C 35 C Relative Humidity 30 60 Power frequency 50 or 60Hz magnetic eld IEC 61000 4 8 30A m 50 or 60Hz 30A m 50 or 60Hz Powerfrequency magnetic eldsshouldbeatlevels characteristicof atypical locationinatypicalcommercial orhospitalenvironment Electrical fast tran...

Page 10: ...ed by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures object...

Page 11: ...399 8 30 am 5 00 pm EST Monday Friday or contact us at info stareliteinc com Our customer service will be happy to assist you 2021 Sunbeam Products Inc All rights reserved Distributed by Star Elite Inc Montreal Canada H3B 3X9 SE005 080621 Upper Arm Blood Pressure Monitor Model HL858A2 LT Printed in China ...

Reviews: