Warnings
4
Warnings
Federal (U.S.A.) law restricts this device to sale by or on the order of a
physician.
This product should only be used by a physician or by a trained individual
under direct supervision of a physician.
Read instructions before use. The Pump must be used strictly in accordance
with these instructions.
Safe use of this Pump is the primary responsibility of the user. The user is
responsible for monitoring this Pump. Contact Clinical/Technical Support if
Pump appears to be operating incorrectly.
All patients should be given a Patient Manual and instructed to read it
carefully.
The Pump must be used only by the person for whom it is prescribed.
Patients should never perform any function or push any button unless
instructed by their health care professional.
Do not allow the Pump to get wet. If the Pump is immersed in any liquid, it
must be replaced with a new Pump.
Transport and storage conditions: -7ºC to 70ºC (20ºF to 158ºF), 10% to 90%
relative humidity and 500 hPa to 1060 hPa atmospheric pressure.
Operating conditions: 10ºC to 43ºC (50ºF to 110ºF), 10% to 90% relative
humidity and 500 hPa to 1060 hPa atmospheric pressure.
Never attempt to open the Pump case. Only the battery cover may be
removed when changing batteries. If the Pump is dropped, it must be
replaced with a new Pump.
This Pump is not to be used for infusion of blood or blood products.
This Pump is not to be used for infusion of life sustaining medications.
Failure to follow manufacturer’s instructions while replacing batteries may
result in loss of program settings and report data. Dispose of batteries
properly after use.
This Pump does not have an air in-line alarm. A Cassette with an air
elimination filter is available.
Safety hazards such as under infusion may be associated with external radio
frequency interference (RFI) or electromagnetic radiation. Typical equipment
which may generate such radiation include x-ray machines, MRI equipment,
and any other non-shielded electrical equipment.