Troubleshooting
47
Problem
Possible Cause
Corrective Action
Compressor starts but shuts
down sometime later.
Water high-temperature
switch is open.
Check the coolant flow and
temperature. Refer to the
Specifications.
Circuit breaker or fuse is
open.
Reset the Circuit breaker or
replace a fuse. Compare the
electric service with system
specifications. Consult a
SCAI Service Center if the
problem persists.
Component failure in the
power circuit.
Check for an open circuit
breaker or fuse. Reset or
replace if necessary. Check
for a faulty component.
Incorrect current draw.
Measure the current. Check
motor winding resistances. If
checks reveal a failed motor
windings or locked rotor,
consult APD Service Center.
Compressor overload relay
opens.
If water and power checks
indicate the utilities are within
specifications, interlocks may
be faulty. Consult a SCAI
Service Center.
System starts but gas
pressures are abnormally
high or low.
Wrong equalization pressure. Refer to Specifications and
the section on Charging and
Venting. Leak check the
system if the pressure is low.
Gas line couplings are not
fully engaged.
Ensure all Aeroquip couplings
are fully engaged and
torqued.
Gas lines are connected
wrong.
Reconnect lines. See the
Installation section.
Summary of Contents for F-20L
Page 2: ......
Page 10: ...6 This page is intentionally blank...
Page 12: ...8 This page is intentionally blank...
Page 23: ...Specifications 19 Figure 2 F 20L Compressor Front View...
Page 24: ...Specifications 20 Figure 3 F 20L Compressor Dimensions Dimensions are in inches and mm...
Page 34: ...30 This page is intentionally blank...
Page 36: ...32 This page is intentionally blank...
Page 48: ...44 This page is intentionally blank...
Page 53: ...Troubleshooting 49 Figure 19 F 20L Compressor Wiring Diagram...
Page 54: ...Troubleshooting 50 Figure 20 F 20L Compressor Wiring Schematic...