Stryker Visum 300® Ceiling-Mounted Exam Light

Assembly, Operations, and Maintenance Manual

Summary of Contents for Visum 300

Page 1: ...Stryker Visum 300 Ceiling Mounted Exam Light Assembly Operations and Maintenance Manual ...

Page 2: ......

Page 3: ...ecting the Sterilizable Light Handle Cover 16 6 4 1 Cleaning 16 6 4 2 Sterilization 16 6 5 Replacing the Halogen Bulb 17 6 6 Replacing Fuses 19 6 7 Replacing the Transformer 20 6 8 Replacing the Aesthetic Covers 22 6 9 Removing the Light Head 24 6 10 Removing the Extension Arm with the Spring Arm 24 6 11 Removing the Down Tube Assembly 25 6 12 Lubricating the Keeper Clip 25 6 13 Adjusting the Spri...

Page 4: ...ed without any hazard for the patient in case of failure of the light It is not intended for use in operating rooms 1 1 Item Installation Package List 1 Down Tube Assembly see Figure 5 Down Tube Ceiling Cover Protective Cover 2 Extension Arm and Spring Arm Assembly see Figures 1 8 9 Plug C Clip Retaining Ring C Clip Retaining Ring Installation Pliers Light Head Keeper Clip Brake Screw Brake Cover ...

Page 5: ... prior to using this equipment 2 This equipment can only be used by or on the order of a physician 3 Test this equipment prior to any procedures This product was fully tested at the factory before shipment 4 Do not remove or disable components It may cause electric shock 5 Attempt no repairs or adjustments unless specifically instructed to do so in this operating manual 6 Disconnect the unit from ...

Page 6: ...tructions in the literature accompanying the product Denotes temperature limits Denotes humidity limits Denotes usage tips and useful information Denotes compliance to CSA Standard C22 2 60601 1 M90 AS 3200 IEC 60601 IEC 60601 2 41 UL 60601 EN 60601 Denotes the date the equipment was manufactured Denotes the manufacturer of the device Denotes product part number Denotes product serial number Denot...

Page 7: ...heet Metal Guard 7 Elbow 2 Down Tube Flange 8 Spring Arm 3 Sheet Metal Enclosure 9 Power Switch 4 Ceiling Cover 10 Light Head 5 Down Tube 11 Light Handle Assembly 6 Extension Arm REVISIONS REV DESCRIPTION ECO DATE CHECK APPROVED Figure 2 VISUM 300 Light Head Figure 3 Sterilizable Handle ...

Page 8: ...to the ceiling super structure as follows Insert and fasten the threaded rods into the ceiling super structure as shown in Figure 5 Install the upper 3 8 nuts and flat washers onto the threaded rods as shown in Figure 5 Depending on down tube length ensure that the bottom of the washer is the correct height above the bottom of the false ceiling Raise the down tube assembly into the ceiling while g...

Page 9: ... Ceiling Ceiling Cover FALSE CEILING Figure 4 Down Tube Super Structure and False Ceiling H Bottom of False Ceiling Down Tube Height H 110mm 1 25 0 220 800mm 1 2 5 4 3 2 1 Figure 5 Down Tube Assembly Preparation 1 Spring Washer 2 Flat Washer 3 Nut 4 Upper Nut 5 Upper Flat Washer ...

Page 10: ...Washer 2 Junction Box 10 Ceiling Structure 3 Conduit and Fitting Customer Supplied 11 Threaded Rod 4 Conduit Fitting Nut Customer Supplied 12 False Ceiling 5 Transformer Mounting Screw 13 Sheet Metal Guard 6 Transformer Insulating Ring 14 Down Tube Flange 7 Transformer 15 Sheet Metal Enclosure 8 Transformer Clamping Ring 5 6 7 9 8 ...

Page 11: ...llation 1 Conduit Fitting Customer Supplied 6 False Ceiling 2 Down Tube Flange 7 Lower Flat Washer 3 Sheet Metal Guard 8 Lower Nut 4 Upper Nut 9 Spring Washer 5 Upper Flat Washer 10 Sheet Metal Enclosure Customer Supplied 11 Conduit Fitting Nut Customer Supplied 8 ...

Page 12: ...causing serious injuries If the keyed washer is not installed the retaining ring can become loose If this occurs equipment can fall from fixtures causing serious injuries 5 Make sure the suspension is seated firmly 6 Insert the plug into the bottom of the down tube 7 Tighten the Phillips head screw to secure the plug 8 7 6 5 4 3 2 1 Figure 8 Installing the Extension Arm 1 Down Tube Axle 5 Keyed Wa...

Page 13: ...ake cover over the spring arm so that the keeper clip slots in the spring arm and brake cover are aligned 5 Push the end of the light head into the spring arm and install the keeper clip 6 Rotate the brake cover 180 until it is lined up with the hole and install the brake screw 3 4 2 5 1 Figure 9 Attaching the Light Head 1 Keeper Clip 2 Brake Screw 3 End of Light Head 4 Brake Cover 5 Slot ...

Page 14: ...ndle The head itself may be rotated 360 and tilted 180 The spring arm and horizontal arm can be moved separately The spring arm has 90 of vertical movement and 360 of horizontal movement The extension arm also has 360 of horizontal movement You can adjust the positioning of the light as shown in Figure 11 6 3 Removing and Reattaching the Sterilizable Light Handle Cover Perform the following steps ...

Page 15: ...NY NAME HERE ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF INSERT COMPANY NAME HERE IS PROHIBITED A DO NOT SCALE DRAWING Bottom View 360 360 D C B A B C D 1 2 3 4 5 6 7 8 E F E F Visum 300 Install Assembly SHEET 3 OF 3 STRYKER ASSEMBLY INCH TEM A UNLESS OTHERWISE SPECIFIED SCALE 1 48 WEIGHT REV DWG NO C SIZE TITLE NAME DATE COMMENTS Q A MFG APPR ENG APPR CHECKED DRAWN FI...

Page 16: ...ioburden from all surfaces using a soft brush 3 Rinse the handle thoroughly under running lukewarm water flushing water through all passages 4 Dry the handle with a lint free cloth and inspect the handle to make sure it is free of debris 6 4 2 Sterilization The sterilization methods in the following table have been validated Sterilization Method Temperature Exposure Time Gravity displacement steam...

Page 17: ... light has been turned off and allowed to cool remove the optional sterile handle if necessary 2 Unscrew the two thumbscrews at the base of the handle assembly and the pull the handle assembly straight out of the light head 3 Firmly grasp the bulb and pull straight out from the handle assembly to remove it 4 Using the screwdriver provided remove the porcelain bulb holder by removing the two screws...

Page 18: ...the replacement bulb into the socket of the porcelain bulb holder 9 Replace the handle assembly Guide the banana clips into the power receptacles ensuring that the handle assembly is seated against the face glass 10 Secure with the two thumbscrews at the base of the handle assembly Light Handle Light Head Side Side ...

Page 19: ...ding it down the tube Note For short down tubes the removal of the extension arm may be necessary in order to obtain access to the flange 3 Grasp the black plastic fuse holders indicated in the picture to the right with two fingers and pull firmly to remove 4 Check the fuses and replace them if necessary For 115V models use a 1 amp fuse 5 Slide the fuse holders see the sample to the right back int...

Page 20: ...tenance Manual 2 Remove the ceiling cover and the sheet metal enclosure from the ceiling 3 Have an electrician power down the room and disconnect the power cord at the flange 4 Tape up the end of the black cable shown so that the loose wires are not spread out and attach a string to the end of the black cable for pulling the cable back through the down tube later in this procedure Warning An elect...

Page 21: ... when all wiring is disconnected and cable is pulled into the down tube 10 Reinstall the three hex screws that hold the commutator assembly Take special note of the visible hole and screw distance shown to the right so as to avoid inserting the commutator too far into the down tube which would not allow installation of the extension arm 11 Connect the transformer wires as shown in Figure 13 or Fig...

Page 22: ...e 14 Transformer Wiring Diagram 230V Model 6 8 Replacing the Aesthetic Covers Warning Disconnect power from the unit before replacing the aesthetic covers Perform the following steps to replace the aesthetic covers 1 Turn off the light and disconnect power from the unit 2 Gently pry the covers loose using a medium standard edge screwdriver be careful not to damage reusable components ...

Page 23: ...ead procedure in this manual then remove the grounding screw and the brake cover from the end of the spring arm so that you can slide the elbow cap collar completely off 5 Reinstall the covers by reversing the removal process Elbow Cap Cover Grounding Screw 6 Align the cover cut out as shown 7 Make sure covers snap securely in place 8 Replace the grounding screw ...

Page 24: ...ension arm with the spring arm according to the following procedure 6 10 Removing the Extension Arm with the Spring Arm Warning Disconnect power from the unit before removing the extension arm with the spring arm After you have removed the light head it may also be necessary to remove the extension arm with the spring arm for maintenance Perform the following steps to do so 1 Remove the light head...

Page 25: ...embly you can reattach it to the ceiling structure according to the Installing the Down Tube Assembly procedure in this manual 6 12 Lubricating the Keeper Clip Perform the following steps to lubricate the keeper clip 1 Remove the light head according to the Removing the Light Head procedure 2 Check the thickness of the keeper clip 3 If the keeper clip is less than 1 5 mm thick replace it 4 Grease ...

Page 26: ...sert a flathead screwdriver into the spring adjustment hole indicated in the picture to the right then let go of the spring arm 4 If the arm drops turn the adjusting screw counter clockwise to increase the spring force 5 If the arm rises turn the adjusting screw clockwise to decrease the spring force ...

Page 27: ... ON position Halogen bulb needs to be replaced Replace the light s halogen bulb Loose connections Check all terminal connections in the light fixture Note If any of these issues cannot be resolved please contact Technical Support or your Stryker representative AC MAIN 115V Circuit Breaker 2 3 7 6 Figure 13 Electrical Block Diagram ...

Page 28: ...t 14 V at light 4 3 A Max Power Delivered by the 115 V Version 50 W 8 3 Lighting Data Central Illumination 37500 Lux 3483 9 fc Color Rendering Index 92 Color Temperature 3978 K Light Field Diameter 150 mm 6 in Total Irradiance 271 W m2 0 034 hp ft2 8 4 Physical Data Light Head Weight 6 6 lbs 3 kg Spring Arm Max Load 6 6 lbs 3 kg Suspension Arm Lengths Extension Arm 800 mm Spring Arm 720 mm Availab...

Page 29: ...Deformation of the system The VISUM light must be inspected once a year to check for the following Connections and markings Loose parts Keeper clip Fully seated C clips Caution Use only Stryker provided parts Note Service schedules circuit diagrams spare parts lists descriptions adjustment instructions and other documents can be made available if desired In case of any faults or damage or for serv...

Page 30: ...nts that its products shall be free of defects of material and workmanship for a period of two years after date of shipment Stryker will provide all parts and service required to restore equipment under warranty to good working condition which may include shipment of replacement parts and phone service consultation to conduct minor repairs Any modifications to this warranty policy are not valid un...

Page 31: ...Customer Service with questions or concerns Stryker Communications 571 Silveron Blvd Flower Mound TX 75028 Toll Free 866 841 5663 1 972 410 7100 For international service locations refer to the Stryker website at the following URL www stryker com ...

Page 32: ...28 1 866 841 5663 www stryker com Stryker Corporation or its divisions or other corporate affiliated entities own use or have applied for the following trademarks or service marks the Stryker logo All other trademarks are trademarks of their respective owners or holders 2017 01 WCR None ...