Figure 42
Instrument Bar
36
Option 2
– Intramedullary Referencing
>
Attach the 3/8
"
IM Drill to the Universal Driver and create a hole
in the location determined by the preoperative X-rays.
T
ib
ia
l
P
re
p
a
ra
ti
o
n
Figure 43
>
Attach the T-Handle Driver to the 5/16
"
IM Rod and slowly pass
into the canal, ensuring clearance. Remove the 5/16
"
IM Rod and
insert it into the body of the Tibial Alignment Jig IM. The assembly
is then inserted into the canal until the isthmus is engaged.
6541-4-538
3/8" IM Drill
6541-4-800
T-Handle Driver
6541-4-516
5/16" IM Rod
0º slope 6541-2-704
3º slope 6541-2-705
Tibial Adjustment Housing
6541-2-600
Tibial Alignment Jig IM
Right 6541-2-700
Left 6541-2-701
Tibial Resection Guide
6541-2-609
Tibial Alignment Ankle Clamp EM
6541-2-611
Tibial Alignment Proximal Rod EM
6541-2-610
Tibial Alignment Distal Assembly EM
6541-4-801
Universal Driver