background image

Table of Contents

Portable Warmer Operation

WARNINGS

2

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

INTRODUCTION

3

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

SPECIFICATIONS

3

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

WARNING / CAUTION / NOTE DEFINITION

3

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

CLEANING

3

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

PRE–OPERATION SET–UP

4

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

OPERATING WARMER

5–7

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

OVERLAY SPECIFICATIONS

8

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

PORTABLE WARMER SAFETY FEATURES

9

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

PORTABLE WARMER SUPPORT POST

10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

USING THE I.V. POLE ATTACHMENT SYSTEM

10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

Troubleshooting  Guide

11

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

Fuse Replacement

11

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

Filter Replacement

12

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

Assembly Drawings and Parts Lists

BASE ASSEMBLY

13

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

HOSE ASSEMBLY

14

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

SHELL AND HOSE ASSEMBLY

15

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

Summary of Contents for Thermaltek 300 Plus

Page 1: ...Thermaltek 300 Plus Portable Warmer OPERATIONS MAINTENANCE MANUAL For Parts or Technical Assistance 1 800 327 0770 IMPORTANT Keep manual with warmer at all times ...

Page 2: ...RE OPERATION SET UP 4 OPERATING WARMER 5 7 OVERLAY SPECIFICATIONS 8 PORTABLE WARMER SAFETY FEATURES 9 PORTABLE WARMER SUPPORT POST 10 USING THE I V POLE ATTACHMENT SYSTEM 10 Troubleshooting Guide 11 Fuse Replacement 11 Filter Replacement 12 Assembly Drawings and Parts Lists BASE ASSEMBLY 13 HOSE ASSEMBLY 14 SHELL AND HOSE ASSEMBLY 15 ...

Page 3: ...her warming blanket may result in thermal injury S The Thermaltek overlay must only be used with the Thermaltek Plus Portable Warmer Using a Ther maltek overlay with any other warming unit may result in thermal injury S The temperature settings on the Thermaltek 300 Plus High 110_ 5_ F 43_ 3_ C Medium 100_ 5_ F 38_ 3_ C Low 90_ 5_ F 32_ 3_ C are average temperatures at various points around the pa...

Page 4: ...UTION and NOTE carry special meanings and should be carefully reviewed WARNING The personal safety of the patient or user may be involved Disregarding this information could result in injury to the patient or user CAUTION These instructions point out special procedures or precautions that must be followed to avoid damaging the equipment NOTE This provides special information to make maintenance ea...

Page 5: ...rtments for warmer overlays Slip the top loop over the top of the base tube and secure the bag to the bottom of the base tube with the velcro strap WARNING The 300 Plus Portable Warmer is equipped with a hospital grade plug for protection against shock hazard It must be plugged directly into a properly grounded three prong receptacle Grounding reliability can be achieved only when a hospital grade...

Page 6: ...table Warmer is equipped with a step down feature that automatically lowers the warmer temperature from high to medium and sounds a brief alarm 3 beeps after 20 minutes of continuous operation see page 9 AMBIENT light indicates room temperature air is being circulated To circulate ambient air press and release the B key until the AMBIENT light comes on Press OFF STANDBY to stop warmer operation Th...

Page 7: ...rlay and the patient WARNING S When using the full body and lower body overlays the urethane hose connector must be positioned so that the hose as it leaves the blanket is parallel to the center tubes of the overlay S When using the upper body overlay the urethane hose connector must be positioned so that the hose as it leaves the blanket is perpendicular to the center tubes of the overlay S Posit...

Page 8: ...ase occlusive or diabetic Medium Low Cardiac Output Medium Totally Immobilized Anesthetized Medium Marginal Cutaneous Perfusion Medium WARNING Monitor the patient s core axillary or skin temperature and skin condition at a minimum of 15 minute intervals during the warming therapy Reduce air temperature or discontinue therapy when the therapeutic goal is reached if vital sign instability occurs or ...

Page 9: ...r and 2 ethylene gas 2 ethylene gas is less than the percentage in the air we breathe Full Body Overlay 92 long x 55 wide Upper Body Overlay 90 long x 30 wide Lower Body Overlay 58 long x 55 wide OVERLAY PART NUMBERS PACU FB Full Body Part Number 300 50 00 OR UB Upper Body Part Number 300 52 00 OR LB Lower Body Part Number 300 54 00 Overlay Storage Bag 300 68 10 ...

Page 10: ...ng the problem reset the desired temperature and air flow settings AUTOMATIC TEMPERATURE STEP DOWN The Thermaltek 300 Plus Portable Warmer is equipped with Automatic Temperature Step Down a feature that automatically lowers the warmer temperature from high to medium and sounds a brief alarm 3 beeps after 20 minutes of continuous operation The warmer will remain on the medium setting until another ...

Page 11: ...e the Plus Por table Warmer only has one support post it al lows flexibility in placement 3 The support post allows the warmer to be placed in any standard 1 2 I V pole socket on a stretcher or bed Special adaptors that allow the warmer to rest in larger I V sockets 5 8 1 may also be purchased USING THE I V POLE ATTACHMENT SYSTEM A B C An I V pole attachment system is included on all Plus Portable...

Page 12: ...erlay disconnected from hose Reattach overlay Reset warmer temperature and air flow set tings High alarm sounds repeatedly af ter all obstructions are resolved Temperature sensor possibly de fective Discontinue use of warmer Contact Stryker representative Warmer shuts off no alarm sounds repeatedly Temperature sensor possibly de fective Discontinue use of warmer Contact Stryker representative Low ...

Page 13: ...B to the inlet shell C The nylon washers and stand offs D will come loose when the cap screws are removed Retain them for reuse 2 Remove the inner inlet screen E and the air filter F Place the replacement filter on top of the outer inlet screen G and replace the inner screen E standoffs and washers D and cap screws A 3 Replace the filter every 4 6 months or if the filter appears soiled ...

Page 14: ...Assembly part number 300 1 10 Item Part No Part Name Qty A 300 1 52 Base Tube Weldment 1 B 37 199 Tube Closure 1 C 37 81 Tube Plug 1 D 300 1 41 Star Base w Casters 1 E 390 38 5 Support Plate 1 F 3 82 Hex Hd Cap Screw 1 ...

Page 15: ...14 Hose Assembly Assembly part number 300 32 10 Item Part No Part Name Qty A 300 32 40 Hose Attachment 1 B 300 32 41 Hose Sleeve Assembly 1 C 300 32 45 Hose End Connector 1 ...

Page 16: ... 30 Hole Plug 1 N 921 1 252 Label Serial No 1 D 300 30 10 Shell Assembly 1 P 1030 34 38 Cap Strap 1 E page 14 Hose Assembly 1 R 300 90 8 Label Caution Hose 1 F 300 32 43 Outlet Cap 1 S 300 90 18 Monitor Patient Tag 1 G 300 90 1 Ops Maint Manual 1 T 300 90 19 Label Monitor Patient 1 H 300 90 2 Label Serial No U L 1 U 58 14 Cable Tie 1 J 300 90 3 Label Display 1 W 300 90 20 Label Step Down 1 K 300 9...

Page 17: ...DH 12 94 300 90 1 REV D 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...

Reviews: