Instructions For Use

Ref 6205

System 6 

Dual Trigger Rotary Handpiece

Summary of Contents for System 6 6205

Page 1: ...2008 01 6205 001 700 Rev A www stryker com Instructions For Use ref 6205 System 6 Dual Trigger Rotary Handpiece ...

Page 2: ...ssories unless other wise specified Other accessories may re sult in increased electromagnetic emissions or decreased electromagnetic immunity of the system DO NOT modify any component or accessory Failure to comply may result in patient and or health care staff injury Important Information The words WARNING CAUTION and NOTE have special meaning and should be reviewed WARNING Disregarding WARNING ...

Page 3: ...ipment like the handpiece Install and place the handpiece into service according to the EMC information in this manual Portable and mobile RF communications equipment can affect the function of the handpiece ALWAYS place the handpiece in the safe mode while the handpiece is idle before installing or removing any accessory or when passing the handpiece to an other person Failure to comply may resul...

Page 4: ...REAM slow or DRILL fast rotate the knob to the appropriate position as required Accessory Interface Connector provides power and communication for future accessories Symbol Definitions FORWARD Fully depress the F R Safe control to lock the reverse trigger The forward trigger is functional REVERSE Press trigger to operate in the counterclock wise direction REAM For slow speed rotate the shifter kno...

Page 5: ...ezing the trigger Slide the F R Safe control to the safe position until you are ready to use the handpiece 2 Slide the F R Safe control to the safe position To install an attachment align the driveshaft of the attachment with the handpiece spindle If the attach ment has notches also align the notches with the tabs in the sleeve of the handpiece see figure 1 1 2 Once properly aligned insert the att...

Page 6: ...e F R Safe control to the safe position when you are finished operating the handpiece 1 2 CAUTIONS DO NOT stall the handpiece Failure to comply may damage the electric motor and or battery pack If the handpiece jams release the trigger immediately Remove any obstructions before continuing the pro cedure If any power loss is experienced while using a hand piece ALWAYS replace the battery pack with ...

Page 7: ...he battery pack on the charger Replace the battery pack if required See the instructions supplied with the battery charger Attachment will not fit into the handpiece Debris is on the attachment or inside the front end of the handpiece Clean the attachment and or handpiece with a small brush with stiff non metallic bristles Attachment is damaged Return the attachment for repair Handpiece is damaged...

Page 8: ...ng properly Ensure that there are no loose or missing components Check all moving parts for free movement Be alert for unusual sounds or vibrations and note the operating speed PROBLEM CAUSE ACTION Handpiece has become noisy and vibrates Drivetrain is malfunctioning Return the handpiece for repair Sporadic electrical interference is experienced Electrical noise is present Turn off all electrical e...

Page 9: ... Handpiece and Attachments Remove the battery pack and attachment from the handpiece Using a brush with stiff non metallic bristles and hospital enzymatic cleaner scrub the debris from the handpiece and attachments Pay special attention to crevices and other hard to reach areas such as seams joints and details around the retainer trigger and connector areas Use a bottle brush to clean the cannula ...

Page 10: ...posure time 8 minute minimum dry time To Sterilize Battery Packs See the care instructions supplied with the battery packs To Sterilize Handpieces and Attachments Sterilization Recommendations ETO 100 ETO 120 135 F 49 57 C Wrapped in an instrument tray or fully perforated sterilization box 2 hour 30 minute exposure time 8 hour minimum aeration time 250 F Gravity Gravity displacement sterilizer 250...

Page 11: ...Intermittent Operation 1 minute on 4 minutes off 3 times with 3 hour rest Approval CSA International CAN CSA C22 2 No 601 1 M90 UL 60601 1 IEC 60601 1 Equipment Type Type BF Applied Part Power Supply Internally Powered 9 6 V Enclosure Protection IPX0 Ordinary Equipment Environmental Conditions Operation Storage and Transportation Temperature Relative Humidity Atmospheric Pressure Specifications ar...

Page 12: ...ssions CISPR 11 Group 1 The System 6 handpiece uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The System 6 handpiece is suitable for use in all establishments includ ing domestic establishments and those directly connected to the public low voltage power ...

Page 13: ...han the recommended sep aration distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 67 P d 1 67 P 80 MHz to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Interference ma...

Page 14: ...erial the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines n a n a Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode n a n a Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle n a 40 UT 60 dip in UT for 5 cycles n a 7...

Page 15: ... output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 n a 0 12 0 23 0 1 n a 0 37 0 74 1 n a 1 17 2 33 10 n a 3 70 7 37 100 n a 11 70 23 30 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable t...

Page 16: ...ts 4100 E Milham Kalamazoo Michigan USA 49001 1 269 323 7700 1 800 253 3210 Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France ES DE FR IT NL JA ZH KO SV DA FI PT NO PL EL 6205 001 710 6205 001 720 6205 001 730 6205 001 750 ...