![Stryker System 6 6203-000-000 Instructions For Use Manual Download Page 16](http://html.mh-extra.com/html/stryker/system-6-6203-000-000/system-6-6203-000-000_instructions-for-use-manual_1386046016.webp)
6203-001-700 Rev-E
EN
www.stryker.com
15
Recommended separation distances between portable and mobile RF communications equipment and the
System 6 Single Trigger Rotary Handpiece (REF 6203-000-000)
The System 6 Single Trigger Rotary Handpiece (REF 6203-000-000) is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of the System 6 Single
Trigger Rotary Handpiece (REF 6203-000-000) can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the System 6
Single Trigger Rotary Handpiece (REF 6203-000-000) as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
d
= 1.2√P
800 MHz to 2,5 GHz
d
= 2.3√P
0,01
N/A
0.12
0.23
0,1
N/A
0.38
0.73
1
N/A
1.2
2.3
10
N/A
3.8
7.3
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.