PROBLEM
CAUSE
ACTION
Motor runs but cutting
accessory does not move.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Attachment is not fully seated.
Remove and insert the attachment. Ensure the
attachment is fully seated.
Shifter knob is positioned between
the DRILL and REAM selections.
Position shifter knob to either the DRILL or
REAM position.
Battery pack becomes
unusually hot during use.
Circuitry is malfunctioning.
Check the battery pack on the charger.
Replace the battery pack if required. See the
instructions for use supplied with the battery
charger.
Attachment will not fit into
the handpiece.
Debris is on the attachment
or inside the front end of the
handpiece.
Clean the attachment and/or handpiece with a
small brush with stiff, non-metallic bristles.
Attachment is damaged.
Return the attachment for repair.
Handpiece is damaged.
Return the handpiece for repair.
Attachment and/or cutting
accessory wobbles in
handpiece.
Cutting accessory or wire/pin is
bent, extends too far from the distal
end of attachment, is the wrong
size, or is not properly centered in
the attachment.
Reinsert the cutting accessory, wire or pin.
If wobble persists, return the handpiece and
attachment for repair.
A reamer is installed and the
handpiece is in the DRILL mode.
ALWAYS operate a reamer in the REAM mode.
Handpiece has become
noisy and vibrates.
Drivetrain is malfunctioning.
Return the handpiece for repair.
Sporadic electrical
interference is
experienced.
Electrical noise is present.
Turn off all electrical equipment not in use in
the operating room.
Relocate electrical equipment; increase spatial
distance.
Plug operating room equipment into different
operating room outlets.