Sterilization Recommendations
WARNINGS:
• Handpieces and batteries must be cleaned and
sterilized before first and every use.
• Remove the sternum blade guard, blade and battery
from the handpiece prior to sterilization.
To obtain optimum performance from the instruments
and help prevent damage to the instruments, it
is essential that one of the following sterilization
procedures be performed.
“Flash” Autoclave:
• Gravity displacement sterilizer
• 270-272°F (132-134°C)
• Unwrapped in an instrument tray
• 10-minute minimum exposure time
Hi Vac:
• Pre-vacuumed sterilizer
• 270-272°F (132-134°C)
• Wrapped or unwrapped
• 4-minute minimum exposure time
• 8-minute minimum dry time
Battery Packs
Refer to the
Sterilization Recommendations
provided
with the battery packs.
ETO:
• 100% ETO
• 120-135°F (49-57°C)
• Wrapped in an instrument tray or fully perforated
sterilization box
• 2-hour 30-minute exposure time, 8-hour minimum
aeration time
250°F Gravity:
• Gravity displacement sterilizer
• 250-254°F (121-123°C)
• Wrapped in an instrument tray or fully perforated
sterilization box
• 50-minute exposure time
• 8-minute minimum drying time
270°F Gravity:
• Gravity displacement sterilizer
• 270-272°F (132-134°C)
• Wrapped in an instrument tray or fully perforated
sterilization box
• 35-minute minimum exposure time
• 8-minute minimum dry time
Sterilization Recommendations
Validation based on AAMI protocol.
Handpiece and Sternum Blade Guard