92
4-0003045 Rev. G
The video image appears
too dark.
The Brightness value is set
too low.
1.
In the VPI Options menu, check the Image
Quality value for Brightness.
2.
If the value is less than 175, increase the
value.
The video image appears
too red or blue.
The Red or Blue Saturation
value is set too high.
1.
In the Options menu, check the Image
Quality value for the Red or Blue
Saturation.
2.
If the Red value is greater than 768 or the
Blue value is greater than 1024, decrease
the value.
Image quality is poor due
to poor color or brightness.
Image quality settings have
been altered significantly.
▪
Reset the Image Quality settings by selecting the
Default Profile from the user options menu.
Other equipment near the
SPY-PHI imager and/or
PINPOINT Endoscopic
Fluorescence Imaging
System seems to
malfunction when the
system is powered on, but
works normally when the
system is powered off.
Radio frequency
interference may be
occurring between the
imaging system and the
malfunctioning equipment.
1.
Check to see if any of the following resolve
the problem:
▪
Reorient or relocate the imaging system or
the equipment receiving the interference.
▪
Increase the separation between the
imaging system and the equipment
receiving the interference.
▪
Connect the imaging system to a power
outlet on a different circuit from that to
which the other equipment is connected.
2.
If the imaging system becomes
unresponsive, and does not resume normal
function after turning off and back on, stop
using the system and contact a qualified
NOVADAQ service representative.
1.
Disconnect the VPI power cord from the wall receptacle.
2.
Carefully remove the fuse cover with the fuse located next to the three-prong
power connector on the rear panel of the VPI.
3.
Replace the fuse (Littelfuse, 021806.3HXP, 5x20mm, T6.3A L 250VAC) with
the same model or a listed fuse with the same ratings.
4.
Re-install the fuse cover.
5.
If the VPI fails to operate properly again, contact a qualified NOVADAQ
service representative for repair.
Summary of Contents for SPY PHI HH9000
Page 1: ......
Page 6: ...iv 4 0003045 Rev G This page intentionally left blank...
Page 8: ...2 4 0003045 Rev G This page intentionally left blank...
Page 36: ...30 4 0003045 Rev G This page intentionally left blank...
Page 58: ...52 4 0003045 Rev G This page intentionally left blank...
Page 102: ...96 4 0003045 Rev G This page intentionally left blank...
Page 106: ...100 4 0003045 Rev G This page intentionally left blank...
Page 110: ...104 4 0003045 Rev G This page intentionally left blank...
Page 111: ......
Page 112: ......