SteP 3
4
Secur-Fit
®
ADVANceD SurgicAL tecHNique
PrePAriNg tHe FeMOrAL cANAL
Figure 5
A clear-out hole has been
placed on the back of the
Box Osteotome that allows
bone chips to be removed
using a curette or other Or
instruments in an action that
does not require a force to
be directed at the cutting
surface.
tip
To help ensure proper final
orientation of the stem, lateral
bias during preparation for the
implant is preferred. Retraction of
the gluteus medius and removal of
a portion of the lateral neck and/
or piriformis fossa will permit true
axial introduction of the instruments
and implant. A Box Osteotome is
useful in removing bone from this
area. Connect the Box Osteotome
attachment to the Modular Handle.
Use a mallet to strike the end of
the Modular Handle with careful
controlled blows to remove remnants
of the superior femoral neck and
open the medullary canal.
instruments
Modular Box Osteotome
1601-1210
Orthonomic Modular Handle
1020-2900
Orthonomic t-Handle
1101-2200