Calcar planing creates a final resection
plane and angle to optimize stem fit.
There are two calcar planers in the set
(standard, large). Select the size that
fits the cut surface of the neck. Each
calcar planer includes a spring-loaded
alignment tab that provides controlled
alignment and engagement with the
broach.
Proper insertion depth of
the broach in the canal is
achieved when it seats tightly
within the canal based on
visual and auditory clues. the
surgeon’s clues to confirm
implant fixation include
increased pitch of sound
with blows to broach handle
and increased resistance to
advancement. reliance only
on the neck cut may lead to
improper sizing, inadequate
component fixation, and
femoral fracture.
cAutiON
Figure 10
9
OPtiONAL SteP: cALcAr PLANiNg
Initiate power prior to contacting the
femur and slowly advance the Calcar
Planer toward the broach. Align the
tab within the broach body and press
down to plane the calcar to desired
level. Failure to operate the Calcar
Planer in accordance with these
instructions may result in damage
to the femur.
Secur-Fit Advanced calcar Planer,
Standard
1601-1400
Secur-Fit Advanced calcar Planer,
Large
1601-1420