Replacing Components and Modules
61
EN
6.16
Uninstalling/Installing Tube Detection
DANGER!
Disconnect device from power supply.
1. Open the device as described in chapter 3.6 Opening the Device .
2. Remove device front panel as described in chapter 6.14 Uninstalling/Install-
3. Remove tube connection as described in chapter 6.15 Uninstalling/Installing
4. Uninstall the MIO board MIO LCD as described in chapter 6.8 Uninstalling/In-
5. Remove adhesive copper tape from device front (see Fig. 6-43 Remove face-
6. Carefully press the six catches (plastic) into the frame so that the faceplate
(plastic) can be removed from the device front (see Fig. 6-43 Remove face-
plate, page 65).
7. Remove the light sensor from the faceplate (see Fig. 6-40 Position of touch
8. Remove the ST6 plug/cable leading to the tube detection
(4)
from the IMO
Fig. 6-35 Position of tube detection
(1)
Micro switch
(2)
Tube detection holder
(3)
1x fastening screw
(4)
ST6 plug/cable leading to MIO
board
(4)
(3)
(2)
(1)
Fig. 6-36 Adjusting the trigger sensitivi-
ty
(5)
Plunger
(6)
Micro switch
(7)
Locknut screw
(8)
Setting screw
(6)
(7)
(5)
(8)