Replacing Components and Modules
49
EN
6.8
Uninstalling/Installing MIO/Board MIO LCD
DANGER!
Disconnect device from power supply.
Uninstalling/installing MIO board
1. Open the device as described in chapter 3.6 Opening the Device .
2. Remove device front panel as described in chapter 6.14 Uninstalling/Install-
3. Remove fluid sensor as described in chapter 6.13 Uninstalling/Installing Fluid
4. Remove the ST1 plug
(3)
and ST8 plug
(6)
from the MIO board by pulling to-
wards side.
5. Remove the ST5 plug
(5)
, ST6 plug
(4)
, and ST10 plug
(7)
from the rear MIO
board by pulling towards side.
6. Remove MIO LCD board by pulling up from the board bracket.
7. Remove the fastening nuts
(1)
and lift out the MIO board.
8. Follow these steps in reverse for assembly.
Perform the following tests/calibrations after installing the new component:
Fig. 6-16 Position of MIO/MIO LCD board
(1)
4x fastening screws
(2)
MIO board
(3)
Plug ST1/cable leading to touch
display
(4)
ST6 plug/cable leading to HPL
board
(5)
Plug ST5/cable leading to light sen-
sor or tube detection
(6)
Plug ST8/flat ribbon communica-
tion cable
(7)
Plug ST10/cable leading to MIO
LCD board
(1)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
A
Fig. 6-17 Detail A: MIO LCD board
(9)
Flat ribbon cable LCD
(10)
ST2 plug latching
(11)
ST2 plug
(12)
MIO LCD board
(13)
Plug ST3/cable leading to MIO
board
(14)
Board bracket
(9)
(10)
(11)
(12)
(14)
A
(13)