Replacing Components and Modules
59
EN
board (see Fig. 6-20 Detail A: Power switch/NET board, position
(5)
).
8. Remove the ST8 plug/cable
(4)
from the IMO board.
9. Lift the device front panel
(1)
from the device case.
10. Follow these steps in reverse for assembly.
The following tests must be performed after installing a new device front:
• Basic function test
• Safety test (see chapter 4.1 Safety Test)
The device front panel is to be removed when replacing the MIO or HPL board or
the touch display, the tube connection and the tube detection to ensure proper
access to these components.
6.15
Uninstalling/Installing Tube Connection
Fig. 6-32 Position of tube connection
(1)
Snap lock
(2)
Patient pressure measurement
connection
(3)
Tube connection
(4)
HPL board (tube heater)
(5)
Tube detection
(1)
(2)
(3)
(4)
(5)
Fig. 6-33 Tube connection
(6)
Tube connection brass ring
(7)
MIO board
(6)
(7)