Service Menu
37
EN
5.14
Deleting Contamination
This menu option does not have to be selected in order to calibrate the device but
serves only to cancel a contamination alarm after a device contamination or after
testing the contamination alarm. Please note: the device must be restarted after
deleting the contamination alarm.
WARNING!
Canceling a contamination alarm is only permissible if this alarm was triggered
by an erroneous manipulation or during service or maintenance tasks (e.g., by
pulling a plug or during cleaning). The contamination alarm is not canceled until
after the device has been restarted. If the contamination alarm was triggered
while using the device with a patient, it is not permissible only to cancel the con-
tamination alarm. In this case, the entire LPU must be replaced.