Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 9 of 61
PIP - Exterior Physical Inspection/Cleaning Paddles
To perform an exterior physical inspection:
1. Inspect the LIFEPAK 20 exterior for the following: Damage, excessive wear, improper mechanical function, and damaged
connectors.
2. Pick up and turn over the device and listen for loose or rattling hardware. Locate any loose or rattling hardware, and then tighten or
replace it.
3. Inspect the rubber feet on the underside of the lower enclosure. Reinstall or replace rubber feet as necessary.
4. Inspect the Therapy, ECG, SpO2 (if equipped), DB-9, DB-15, AED door, and IrDA connectors for damage and or cracks.
5. Inspect the keypads and overlays for damage, cracks, and separations.
6. Check all other accessory cables, Paddles, ECG, SpO2 sensors, and related items for expiration dates, general condition, and
suitability for use.
7. Inspect carrying strap and mounts (if the device is equipped with them).
To perform Cleaning Paddles:
1. Disconnect the adult paddle plate from the paddle assembly.
2. Clean the spring contact of the adult paddle with alcohol.
3. Clean the pediatric electrode surface with alcohol.
4. Reattach the adult paddle plate to the paddle assembly.