Stryker LF2019 Instructions For Use Manual Download Page 5

Remanufactured LigaSure Exact Dissector, Without Nano-coating 

Page 5 of 9 

 

Do not divide tissue before the seal cycle is complete as this may result in improper sealing. 

 

Do not pull the cutting trigger while the jaws are open as injury to the patient or surgical team may occur. 

 

Conductive fluids (e.g., blood or saline) in direct contact with or in close proximity to the instrument may carry electrical 

current or heat, which may cause unintended burns to the patient. Aspirate fluid from around the instrument jaws before 

activating the instrument. 

 

Keep the external surface of the instrument jaws away from adjacent tissue while activating the LigaSure system or 

unintended injury may result. 

 

During a seal cycle, energy is applied to the tissue between the instrument jaws. This energy may cause water to be converted 

into steam. The thermal energy of steam may cause unintended injury to tissues in close proximity to the jaws. Care should be 

taken in surgical procedures occurring in confined spaces in anticipation of this possibility. 

 

Use caution when grasping, manipulating, sealing, and dividing large tissue bundles. 

 

Do not attempt to seal or cut over clips or staples as incomplete seals/damage to the cutting blade will occur. Contact between 

an active electrode and any metal objects may result in alternate site burns or incomplete seals. 

 

The surfaces of the jaws may remain hot enough to cause burns after the RF current is deactivated. 

 

Inadvertent activation or movement of the activated instrument outside of the field of vision may result in injury to the patient 

or surgical team. 

 

Do not activate the LigaSure system in an open-circuit condition. Activate the system only when the instrument is in direct 

contact with the target tissue to lessen the possibility of unintended burns. 

 

Do not activate the instrument while instrument jaws are in contact with, or in close proximity to, other instruments including 

metal cannulas, as localized burns to the patient or physician may occur. 

 

Eliminate tension on the tissue when sealing and cutting to ensure proper function and reduce bleeding. 

 

Precaution

 

When intending only to grasp or manipulate tissue, ensure pressure to the ring handles does not cause the second click that 

activates energy delivery. 

 

Keep the instrument jaws clean. Build-up of eschar may reduce the seal and/or cutting effectiveness. Wipe jaw surfaces and 

edges with a wet gauze pad as needed. 

 

Do not overfill the jaws of the instrument with tissue, as this may reduce device performance. 

 

Hand Activation 

Notice

 

Closing the ring handles to the second click activates or reactivates energy delivery if hand-activation is being used. 

 

1.

 

Ensure hand-activation is enabled on the appropriate port. Refer to the generator user’s guide if needed. 

2.

 

Grasp the intended vessel and/or tissue in the center of the jaws. 

3.

 

Close the ring handles until you hear or feel two clicks. At the second click, energy is delivered. 

 

A continuous tone sounds to indicate the activation of RF energy. When the activation cycle is complete, a two-pulsed Seal-

Cycle-Complete tone sounds and RF output ceases. 

 

Notice

 

The surgeon may inspect the seal before cutting the vessel or tissue. After inspecting the seal, the surgeon should create a 

second seal adjacent to the first seal before cutting, as described below. 

 

A tone with multiple pulses indicates that the seal cycle was not completed. Refer to the Troubleshooting section for 

possible causes and corrective actions. Do not cut tissue until you have verified that there is an adequate seal.

 

Summary of Contents for LF2019

Page 1: ...entified ISO 15223 1 2016 2492 Batch code Indicates the manufacturer s batch code so that the batch or lot can be identified ISO 15223 1 2016 1641 Consult instructions for use Indicates the need for the user to consult the instructions for use ISO 15223 1 2016 1051 Do not re use Indicates a medical device that is intended for one use or for use on a single patient during a single procedure ISO 152...

Page 2: ... and division of vessels lymphatics and tissue bundles 2 3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures Do not use the LigaSure system for these procedures Contraindications for Use The LigaSure system has not been shown to...

Page 3: ...ach side arrow shows direction for cutting Cable cord Connector purple and white Precaution The nano coating featured on the original device is not present on this product Adjunctive use of nerve monitoring device is recommended during nerve sparing procedures such as thyroidectomies radical neck dissection and parotidectomies Secondary hemorrhaging after tonsillectomy is a potentially serious adv...

Page 4: ...nerator Refer to the generator user s guide for additional set up information During Surgery The LigaSure Sealer Divider can be used during surgery both to manipulate and dissect tissue and to seal and cut vessels and tissue bundles Instructs for use of the instrument during a procedure are provided in this section Warning Avoid placing fingers between the jaws or ring handles Injury to the use ma...

Page 5: ...the system only when the instrument is in direct contact with the target tissue to lessen the possibility of unintended burns Do not activate the instrument while instrument jaws are in contact with or in close proximity to other instruments including metal cannulas as localized burns to the patient or physician may occur Eliminate tension on the tissue when sealing and cutting to ensure proper fu...

Page 6: ...ssel or tissue After inspecting the seal the surgeon should create a second seal adjacent to the first seal before cutting as described below A tone with multiple pulses indicates that the seal cycle was not completed Refer to the Troubleshooting section for possible causes and corrective actions Do not cut tissue until you have verified that there is an adequate seal 6 To seal adjacent tissue ove...

Page 7: ... The user is grasping thin tissue or not enough tissue open the jaws and confirm that a sufficient amount of tissue is inside the jaws If necessary increase the thickness of the tissue that is grasped and reactivate the seal cycle Too much tissue between the jaws The user is grasping too much tissue open the jaws reduce the amount of tissue tat is grasped and reactivate the seal cycle Activating o...

Page 8: ...of one year from the date of purchase General Warranty Terms Applicable to All Products TO THE FULLEST EXTENT PERMITTED BY LAW THE EXPRESS WARRANTY SET FORTH HEREIN IS THE ONLY WARRANTY APPLICABLE TO THE PRODUCTS AND IS EXPRESSLY IN LIEU OF ANY OTHER WARRANTY BY STRYKER EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN ...

Page 9: ...he initial warranty period has expired by the time the product is repaired or replaced for thirty 30 days after delivery of the repaired or replaced product When a product or component is replaced the item provided in replacement will be the customer s property and the replaced item will be Stryker s property If a refund is provided by Stryker the product for which the refund is provided must be r...

Reviews: